Kusam-meco KM 175D Operation Manual Download Page 10

®

®

17

06

ELECTRICAL SPECIFICATIONS :

Accuracy is ± (% of reading number of digits) or otherwise 

o

o

specified, at 23 C ± 5 C.

Maximum Crest Factor <2.5:1 at full scale & <5:1 at half scale or 
otherwise  specified,  and  with  frequency  components  within  the 
specified  frequency  components  within  the  specified  frequency 
bandwidth for non-sinusoidal waveforms.

Range

TM

AMPTIP  CLAMP-ON AC CURRENT

1) 2) 3) 4)

   Accuracy

     Resolution

50Hz ~ 60Hz

±(1.5%rdg + 5dgts)

0.1 A

60.00  A

1)

 

Induced error from adjacent current-carrying conductor :

 < 0.06A/A

2)

 Induced error from ACV measurement <0.60A / kV @ 50 / 60Hz

3)

 Add 10d to the specified accuracy @ < 6A

4)

 Induced non-zero residual while beeper turns on : <20d

REGULAR CLAMP-ON AC CURRENT

1)

 

Induced error from adjacent current-carrying conductor :

 < 0.06A/A

2)

 Induced error from ACV measurement <0.60A / kV @ 50 / 60Hz.

3)

 Specified accuracy is for measurements made at the jaw center.  When the 

   conductor is not positioned at the jaw center,add 2% to specified accuracy for position errors.

4)

 Induced non-zero residual while beeper turns on : < 20d

5)

 Add 10d to specified accuracy @ < 6A.

1) 2) 3) 4)

   Accuracy

±(1.8%rdg + 5dgts)

     Resolution

0.01  A

0.1    A

Range

50Hz ~ 100Hz

5)

60.00  A

600.0  A

±(2.0%rdg + 5dgts)

0.01  A

0.1    A

100Hz ~ 400Hz

5)

60.00  A

600.0  A

2. When using Capacitance function, discharge capacitor(s) before

making  any  measurements.  Large  value  capacitors  should  be
discharged through an appropriate resistance load.

TEMPERATURE FUNCTION :

Inputs are made via the test lead terminals COM/+. Press SELECT 
button momentarily to toggle between °C (celsius) &°F (Fahrenheit).

Note :

 Be sure to insert the banana  plug  type-K temperature  bead 

probe  Bkp60  with  correct 

+  -

  polarities.    You  can  also  use  a  plug 

adapter  Bkb32  (Optional  purchase)  with  banana  pins  to  type-K
socket to adapt other type-K standard mini plug temperature probes.

Summary of Contents for KM 175D

Page 1: ...rature 20 C 60 C 80 R H with battery removed o o o o o Temperature Coefficient Nominal 0 15 x specified accuracy C 0 C 18 C or 28 C 40 C or otherwise specified Power Supply Standard 1 5V AAA Battery x...

Page 2: ...52 F 1 0 0 8 C Accuracy1 2 1 0 1 C 1 0 1 5 F 1 0 2 F 1 2 K type thermocouple range accuracy not included Accuracies assume meter interior has the same temperature of the ambient isothermal stage for a...

Page 3: ...ometer Distance Meter Digital Lux Meter Network Cable Tester Power Factor Regulator Earth Resistance Tester Digital Panel Meters DC Power Supplies High Voltage Detector Calibrators Gas Analysers Frequ...

Page 4: ...dent or abnormal conditions of operation or handling KUSAM MECO authorized dealer shall extend this warranty on new and unused products to end user customers only but have no authority to extend a gre...

Page 5: ...WARNING To reduce the risk of fire or electric shock do not expose this product to rain or moisture The meter is intended only for indoor use Keep your hands fingers behind the hand finger barriers o...

Page 6: ...ve conductors But still individual protective equipment must be used if hazardous live parts in the installation where measurement is to be carried out could be accessible CAUTION Disconnect the test...

Page 7: ...range Backlighted LCD display Duo Core to simultaneously measure clamp on Amps A DMM functions VFD V Hz for fundamental V Hz of most Variable Frequency Drives Display Hold Non Contact EF Detection NC...

Page 8: ...abled automatically in this mode Hold feature freezes the display for later view LCD turns on Press the HOLD button momentarily to toggle the hold feature The Complementary Beeper feature can be selec...

Page 9: ...lead terminals COM Application Notes on Flame Sensors The DCmA function is designed especially for HVAC R flame sensor applications The 0 1 mA resolution is useful for identifying the minute current...

Page 10: ...nt 0 60A kV 50 60Hz 3 Specified accuracy is for measurements made at the jaw center When the conductor is not positioned at the jaw center add 2 to specified accuracy for position errors 4 Induced non...

Page 11: ...must be disconnected from the circuit to obtain an accurate measurement reading Capacitance Diode functions Inputs are made through the test lead terminals COM Defaults at Capacitance Press SELECT bu...

Page 12: ...from 2 of range CAUTION Using Resistance Continuity function in a live circuit will produce false results and may damage the meter In many cases the suspected component s must be disconnected from the...

Page 13: ...0 W W Response Time 32ms approx 3 000 V 1 5 rdg 5dgts DIODE TESTER Accuracy Resolution Range 1 mV Test Current 0 3mA typically Open Circuit Voltage 3 5VDC typical Non Contact EF Detection NCV An anten...

Page 14: ...00V CAT III 300V CAT IV V W Hz EF V OFF AUTO H P RMS 8 8 8 8 mFA V VFD P RMS SELECT HOLD COM CAT III 600V CAT IV 300V 2 3 1 4 6 10 mA A AUTO kW Hz 6 1 7 5 Duo The detected Electric Field strength is i...

Page 15: ...much as possible to get the best measuring accuracy For removal press the jaw trigger and remove the jaws from the conductor s CAUTION Adjacent current carrying devices such as transformers motors an...

Reviews: