5.5 Error messages
5.5.1 Error messages in Measurement mode
Error message
Type
Description
Corrective action required
INTL ERR. nn
*
Internal error in
Switch off the power and try again. If the problem
converter operation
persists contact KHRONE MARSHALL service.
NO SIGNAL
N
No signal from
No flow through the primary or Vortex sensor problem.
the vortex sensor
If sensor problem, contact KHRONE MARSHALL service.
LOW FREQ
N
Vortex frequency
Check Flow rate>q min else call KHRONE MARSHALL
Service.
too
low
HIGH FREQ
N
Vortex frequency
Check Flow rate<q max else call KHRONE MARSHALL
Service.
too
high
LOW FLOW
N
Flow rate lower
than minimum flow
Converter will continue to display actual flow rate. However
rate q min
accuracy of measurement may suffer.
HIGH FLOW
N
Actual flow rate
Corrective action depends on application process. If flow
higher than q max
rate exceeds too much further it may damage entire instrument
physically!
INV. CONFIG
F
Configuration date
Check entire configuration again. If error persists-call
in non-volatile memory
KHRONE MARSHALL service.
is not valid.
UC FAIL
F
Micro-Controller
failure
Contact KHRONE MARSHALL Service.
LCDC FAIL
F
LCD Controller IC (s) has
Contact KHRONE MARSHALL Service.
failed (This error is
applicable to users having
display
option)
KEY FAIL
F
Failure of Keys
Contact KHRONE MARSHALL Service.
(applicable to instrument
which have display option)
NOVRAM FAIL
F
Non Volatile Memory
Contact KHRONE MARSHALL Service.
has
failed
LINE INTER
N
Supply interrupted
Press keys to reset this error.
·
This text is displayed for Fatal Errors. The nature of error is beyond the scope of user (failure of IC or other hardware inside converter electronics
etc.). All that can be done is to switch OFF power and then try again. If error message continues, call KHRONE MARSHALL service. internal
error nn = internal error number.
·
Type N indicates non-fatal errors whereas F indicates fatal errors.
·
Measurement stops if any Fatal error is encountered. This means that device makes flow rate equal to zero,
·
Current output to minimum value 4m A & pulse output OFF.
·
When errors are displayed during the measuring mode, 'n Err" (n = number) will appear in the 1st line. n gives the number of momentarily
occurring errors that are displayed alternately with the actual measured value.
·
Error messages disappear when their cause disappears
5.5.2 Error Messages in Programming mode
Error in programming mode can only occur while entering any numerical value. When you enter numerical value outside possible limits you get
message " n.nnn E n" in 1st line and
MIN. VALUE
or
MAX VALVE
or in the 2nd line
(
"MIN VALUE"
if entered value is less than lower limit and
“
MAX VALVE
“ if higher limit is crossed
)
. Note the message indicating permissible limit then press the key to continue.
16
Summary of Contents for VFM 1091
Page 1: ...4 2001 Vortex flowmeter Installation and Operating Instructions VFM 1091 I...
Page 12: ...12...
Page 13: ...13...
Page 30: ...Certificate number of VFM 1091 I Ex FM CSA PTB approvals are pending 30...
Page 31: ...31...
Page 32: ...32...
Page 33: ...33...