OPERATION
5
45
OPTISYS TUR 1060
www.krohne.com
01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en
The device can display from one up to four digits to the right of the decimal place. The default
setting are two decimal places. Four digits right to the decimal point are only possible for
measuring results below 10 NTU/FNU. Each decade upwards decreases the digits right to the
decimal place by one.
If the last or the last two digits right of the decimal place are not stable, you can reduce the
resolution (i.e. the amount of digits right to the decimal place) to hide these digits and get a
stable reading.
5.9.8 Display backlight brightness
Changing a setting and switching to the next item
•
Use the buttons
and
to change the setting.
•
Accept your selection by pressing
.
i
The device switches to the next item of the menu.
The display backlight brightness is of particular interest if multiple devices are located in the
same area. If all of them should have the same clearness of display, then you might have to
adjust the backlight brightness of particular devices. The default setting is 8, ten levels are
available.
5.9.9 Setting the units
Changing a setting and switching to the next item
•
Use the buttons
and
to change the setting.
•
Accept your selection by pressing
.
i
The device switches to the next item of the menu.
The most common unit of the measured value and the default setting is NTU (Nephelometric
Turbidity Units), but FNU (Formazin Nephelometric Units) is also available.