OPERATION
40
OPTISYS TUR 1060
www.krohne.com
01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en
5.9.1 Setting the 4...20 mA output inclusive error level
•
Use the buttons
and
^
to select "CONFIG" and press
.
•
Use
or
to select "4-20".
Use the buttons
and
to set "ON" and press
.
Your task is now to set the lower and the upper turbidity limits:
•
Use the buttons
and
to define the lower turbidity limit which is assigned to the
4 mA output level and press
.
i
The device switches to the next menu item, i.e. the setting of the upper limit.
•
Use the buttons
and
to define the upper turbidity limit which is assigned to the
20 mA output level and press
.
i
The device switches to the next menu item, i.e. "Error level".
INFORMATION!
•
While in calibration or configuration mode, the device has a time-out function; it
automatically returns the device to the normal operation mode ("AUTO") after a 15 minute
period of inactivity.
•
While in the calibration or configuration mode, the relay contacts hold at the last valid
condition and will not change state.
•
You can leave the configuration mode at any time and menu by pressing the button
^
. The
device automatically saves all changes.
•
The order of the following sections and subsections represents the menu topology.