Symptom
Displayed message
Possible cause
Correction
CM-
36dG
CM-
36d
Install the white calibration
plate and appropriate
target mask, and then
perform calibration.
White calibration was
not performed using the
correct procedure.
Perform white calibration
using the provided white
calibration plate and
appropriate target mask.
Install the gloss calibration
plate and appropriate
target mask, and then
perform calibration.
Gloss calibration was
not performed using the
correct procedure.
Perform gloss calibration
using the provided gloss
calibration plate and
appropriate target mask.
Light shielding during 0%
calibration was insufficient.
0% calibration was not
performed using the
correct procedure.
Install the white calibration
plate correctly and perform
correct 0% calibration with
light completely blocked
between the illumination
window and receptor
window.
Light intensity at 100%
calibration was insufficient.
100% calibration was
not performed using the
correct procedure.
Install the white calibration
plate correctly and perform
correct 100% calibration
for the specimen that will
be measured.
A target mask is not
installed. Install the target
mask.
A target mask is not
installed.
Install the appropriate
target mask before
performing measurement.
The UV control coefficient
does not exist. Perform
UV control using the UV
Control Software.
Measurement is not
possible because UV
control has not been
performed for the set
conditions.
Perform UV control for the
set UV conditions from the
UV Control Software that is
provided with SpectraMagic
NX before performing
measurement.
Calibration data has not
been set.
The calibration data for
white calibration has not
been written.
Write the calibration data
to the instrument.
Value exceeds the range
that can be measured by
the instrument.
Color measurement of that
specimen is not possible.
To perform color
measurement, measure
using a different specimen.
The specified operation
cannot be executed
because measurement is in
progress.
The operation cannot
be executed because
measurement is in
progress.
Wait for measurement to
end before executing the
operation.