background image

20

 Connecting the AC Adapter

Notes

  •  To supply AC power to the instrument, always use the AC adapter (AC-A312F) that was supplied with the 

instrument.

 

•  Insert the AC adapter plug all the way.

Operating Procedure

1  Check that the power of both the instrument and 

computer is OFF (LED lamp is not illuminated).

2  Connect the DC output plug of the AC adapter to the DC 

input terminal on the side of the instrument.

3  Connect the AC adapter power plug to a 100 V (50/60 Hz) 

AC outlet.

Notes

  Be sure that the power switch is OFF before inserting or 

removing the DC output plug of the AC adapter.

 Turning the Power ON/OFF

Operating Procedure

Turning the Power ON

1  With the power OFF, press and hold the power key for 

around 1 second.

The power turns ON and the LED lamp above the power key 
illuminates blue.

Notes

  •  After pressing the power key, the LED lamp flashes until 

power ON is completed. While the LED is flashing, the start 
process is in progress. Do not unplug the AC adapter.

Turning the Power OFF

1  Press and hold the power switch for around 3 seconds.

The power will turn OFF. The LED lamp flashes and then turns off.

Notes

  •  After pressing the power key, the LED lamp flashes until 

power OFF is completed. While the LED is flashing, the 
shutdown process is in progress. Do not unplug the AC 
adapter.

Summary of Contents for CM-36d

Page 1: ...Spectrophotometer CM 36d CM 36dG En Instruction Manual Please read before using the instrument...

Page 2: ...s direct current DC This symbol indicates class II protection against electric shock Trademarks Windows is a registered trademark of Microsoft Corporation in the United States and other countries The...

Page 3: ...mble or modify the instrument or the AC adapter Doing so may cause a fire or electric shock The instrument should not be operated if it or the AC adapter is damaged or if smoke or abnormal odor occurs...

Page 4: ...n and close There is the risk of injury Do not use the instrument if the specimen measuring port measurement area is in the line of sight Doing so may result in injury to the eye Take sufficient care...

Page 5: ...tenance or for other reasons be sure to use the packing materials to minimize shock or vibration If the packing materials are lost or damaged contact a KONICA MINOLTA authorized service facility Sampl...

Page 6: ...es and condensation Do not use the instrument in areas where dust smoke or chemical gases are present or in extremely humid environments Do not use the instrument at altitudes higher than 2 000 m Do n...

Page 7: ...target mask inner surface black coated surface with a hand or allow it to become dirty or scratched When the target mask is not in use be sure to place it inside the accessory case CM A286 for storage...

Page 8: ...rature and or humidity may exceed the allowable storage range resulting in malfunction Do not leave the instrument in such places The calibration plate may become discolored if left in a place that is...

Page 9: ...aterials to minimize shock or vibration When sending the instrument in for service package and send the instrument and all accessories Maintenance andInspection To maintain measurement accuracy the in...

Page 10: ...Measurement Procedure 18 Preparation 19 Connecting to a Computer 19 Connecting the AC Adapter 20 Turning the Power ON OFF 20 Installing the Target Mask 21 Mask Detection Function 22 Installing the Zer...

Page 11: ...ng dusty or scratched and to protect it from external light Target Mask CM A291 A292 A293 A294 This is used to change the illumination area specimen measuring port size according to the specimen Each...

Page 12: ...reign substances from entering the instrument when it is stored in a location where there is much dust Cleaning Cloth Used to clean the gloss calibration plate Accessory Case CM A286 This case is used...

Page 13: ...00w Ver 3 2 or later This computer software is used to control the instrument and manage data from a computer Plastic Cell CM A130 optical path length 2 mm CM A131 optical path length 10 mm CM A132 op...

Page 14: ...ionPlate CM A289 USB Cable IF A17 2 m Target Mask SAV CM A294 Target Mask LAV CM A291 Target Mask LMAV CM A292 Computer commercially available Standard accessory Connectable Optional accessory Target...

Page 15: ...able Optional accessory Target Mask MAV CM A293 White Calibration Plate CM A288 Zero Calibration Box CM A290 Accessory Case CM A286 DVD for Color Data Software SpectraMagic NX Professional version Bas...

Page 16: ...asurement specimen white calibration plate or gloss calibration plate Jig mounting screw holes These screw holes are used to mount jigs or other components for fastening the specimen Transmittancespec...

Page 17: ...arts that open and close The lock can be released when you want to slide the cover freely Lift up the knob and rotate it 90 degrees to release the lock and free the cover to be opened and closed Notes...

Page 18: ...s possible to perform measurement on the instrument side Illuminated blue Indicates that measurement is possible Illuminated orange Indicates that calibration has not been completed Not illuminated In...

Page 19: ...asurement values Target Mask Use a blower to blow off dirt or dust on the target mask If the dirt on the outside or inside surface of the target mask does not come off easily remove the target mask fr...

Page 20: ...libration P 23 Install the transmittance zero calibration plate and execute 0 calibration P 26 Executing white calibration 100 calibration WhenusingWAA WavelengthAnalysis Adjustment executeitafterwhit...

Page 21: ...sufficient length Putting tension on the cable may cause connection failure or wire breakage Firmly push in the USB cable connector that matches the shape of the port connection terminal until it can...

Page 22: ...ving the DC output plug of the AC adapter Turning the Power ON OFF Operating Procedure Turning the Power ON 1 With the power OFF press andhold the power key for around1second The power turns ON and th...

Page 23: ...e target mask installed for a prolonged period The target mask may become stuck to the sample holder Procedure 1 Pull the sample holder toward you andholdit so it is opened 2 Pull the currently instal...

Page 24: ...the instructionmanual Notes Incorrect detectionmay occur if the target mask is not installed correctly or if it is dirty or scratched 3 Check that the indicator display matches the size of the install...

Page 25: ...Use of the Zero Calibration Box Do not apply any force to the zero calibration box after it is installed Doing so may cause the zero calibration box to fall off Be careful not to scratch or allow fing...

Page 26: ...de of the white calibrationplate Installing the White Calibration Plate Precautions for Use of the White Calibration Plate Memo White calibration data is used whenperforming white calibrationusing the...

Page 27: ...enit is purchased The gloss calibration plate may become discolored if left exposed to light Therefore make sure to close the cap when the plate is not in use in order to prevent the plate from being...

Page 28: ...install the transmittance zero calibrationplate into the transmittance sample holder For information about installing the transmittance specimenholder refer to the instructionmanual that was provided...

Page 29: ...ffect on the measurement results if the optional transmittance specimenholder CM A96 is installed However in this case it is necessary to performmeasurement with the zero calibration transmittance sam...

Page 30: ...onplate Transmittance specimen chamber Reflectance Measurement P 29 When measuring the reflectance of a film or plate specimen install the specimen into the sample holder before setting it onto the in...

Page 31: ...er toward you andholdit so it is openedin order to protect the sample surface Donot openthe transmittance specimenchamber cover during measurement Doing so may prevent accurate measurement Memo If you...

Page 32: ...the measurement type is approximately 0 0 Although this does not completely match the definition in JIS Z 8722 or other standards it can be used as a relative value for management For setting specime...

Page 33: ...hich contains minute particles accuracy willbe unstable due to precipitation of the particles 3 Close the transmittance specimen chamber cover 4 Use the optional SpectraMagic NX CM S100w color data so...

Page 34: ...facility Regularly scheduled device calibration is required Have regularly scheduled device calibration performed A certain amount of time passed after regularly scheduled device calibration For regu...

Page 35: ...elength correction WAA license has expired The wavelength correction license has expired To renew the wavelength correction license contact a KONICA MINOLTA authorized service facility The light inten...

Page 36: ...sing the correct procedure Install the white calibration plate correctly and perform correct 100 calibration for the specimen that will be measured A target mask is not installed Install the target ma...

Page 37: ...service facility There was an operating failure of light output circuit charging inside the instrument Charging of the light output circuit is not completed The light output circuit is not operating...

Page 38: ...the inside of the integrating sphere Foreign matter or dirt larger than several mm may affect the measurement results Foreign matter and dust inside the integrating sphere can easily scratch the bariu...

Page 39: ...ort to blow these substances off If such substances cannot be removed with the blower performing zero calibration and white calibration may improve performance 17 Has correct calibration data been wri...

Page 40: ...escence reflectance through numerical processing of a measurement object s reflectance under the UV full light source and under the UV cut light source The reflectance of the cut wavelength region for...

Page 41: ...4 Repeatability Colorimetric values Standard deviation within E ab 0 02 Spectral reflectance Standard deviation within 0 1 When a white calibration plate is measured 30 times at 10 second intervals a...

Page 42: ...E measurement Minimum interval between measurements Approx 4 seconds SCI SCE measurement Approx 4 5 seconds SCI SCE GLOSS measurement Approx 4 seconds SCI SCE measurement Sample viewer function Using...

Page 43: ...41 Dimensions Unit mm CM 36dG 4 M3 depth 5...

Page 44: ...42 Unit mm CM 36d 4 M3 depth 5...

Page 45: ...RESULTING FROM THE MISUSE MISHANDLING UNAUTHORIZED MODIFICATION ETC OF THIS PRODUCT OR FOR ANY INDIRECT OR INCIDENTAL DAMAGES INCLUDING BUT NOT LIMITED TO LOSS OF BUSINESS PROFITS INTERRUPTION OF BUS...

Page 46: ...9222 AD2X 11 CBCADA En 2021 KONICA MINOLTA INC...

Reviews: