22
Mask Detection Function
This product detects the size of the target mask that is installed, and automatically changes the instrument
light-receiving lens to the correct position.
Operating Procedure
1 Use the optional software (SpectraMagic NX) and turn
ON the mask detection function.
2 Install the target mask.
The light-receiving lens moves to the correct position
for the size of the detected target mask.
Memo
• For the installation procedure, refer to “Installing the Target
Mask” on P.21 of the instruction manual.
Notes
•
Incorrect detection may occur if the target mask is not
installed correctly, or if it is dirty or scratched.
3 Check that the indicator display matches the size of the
installed mask.
If the sizes do not match, check the following two points and perform the operation again.
• Check that the target mask is installed correctly.
• Check that there is no dirt or scratching on the mounting surface of the target mask. If there is dirt, use a blower
to remove any dust, dirt, and other substances. If the dirt does not come off easily, wipe using a soft cloth
dampened with ethanol.
If the above does not resolve the problem, contact a KONICA MINOLTA-authorized service facility.