Error
No.
Symptom
Item to Check
Corrective Action
Page to
Refer
6
Measure-
ment value
appears as
“------“.
Does data exist?
This appears when there is no data
in measurement value, stored data,
calibration value and target color.
-
Doesn’t color space
become color tempera-
ture?
This appears when color tempera-
ture cannot be converted for display.
Available display range is as follows:
2300 T 20000(K)
|
∆
uv|
<
0.1
34
Doesn’t measurement
value become larger by
user calibration?
This appears when luminance value
exceeds the available display area in
this instrument.
-
Do you set target color
for color difference
measurement?
This appears in color difference dis-
play when target color is not set.
78
7
Measurement
values vary.
Is object to measure
stable?
Keep object to measure stable.
-
Isn't object to measure
of low luminance?
Repeatability of x, y worsens if ob-
ject of low luminance is measured.
It becomes worse especially when
measurement angle is 0.2˚ or 0.1˚.
Also it becomes worse especially
when measurement time is FAST
or Super-FAST. Measure in slower
measurement time.
4
28
Is measurement sync
frequency appropriate
when measuring dis-
play?
Set appropriate measurement sync
frequency for proper measurement.
30
Haven't ambient tem-
perature and humidity
rapidly changed?
Conduct measurement under envi-
ronment free from such fluctuation.
3
8
Strange
measurement
value ap-
pears.
Is objective lens clean? Wipe off with dry and soft cloth or lens
cleaning paper.
106
Has user calibration
been performed cor-
rectly?
Redo user calibration.
64
Haven't you put close-
up lens on?
Select lens type setting correspond-
ing to applied close-up lens.
62
Haven't you put ND
filter on?
Select lens type setting correspond-
ing to applied ND filter.
40
CS-200_E101-118.indd 114
12.7.17 4:52:17 PM
Summary of Contents for CHROMA METER CS-200
Page 21: ...19 Installing CS 200_E001 026 indd 19 12 7 17 4 43 25 PM...
Page 29: ...27 Setting CS 200_E027 060 indd 27 12 7 17 4 46 00 PM...
Page 57: ...55 Setting BACK LIGHT CS 200_E027 060 indd 55 12 7 17 4 46 28 PM...
Page 62: ...60 CS 200_E027 060 indd 60 12 7 17 4 46 32 PM...
Page 63: ...61 Measurement Preparation CS 200_E061 86 indd 61 12 7 17 4 47 50 PM...
Page 88: ...86 CS 200_E061 86 indd 86 12 7 17 4 48 17 PM...
Page 89: ...87 Measurement CS 200_E087 96 indd 87 12 7 17 4 49 12 PM...
Page 98: ...96 CS 200_E087 96 indd 96 12 7 17 4 49 24 PM...
Page 99: ...97 Communication CS 200_E97 100 indd 97 12 7 17 4 51 08 PM...
Page 102: ...100 CS 200_E97 100 indd 100 12 7 17 4 51 09 PM...
Page 103: ...101 Description CS 200_E101 118 indd 101 12 7 17 4 52 11 PM...
Page 124: ...9222 1892 A1 2005 KONICA MINOLTA INC BIIBPK Printed in Japan...