50
S
e
tt
in
g
Setting of Update Method for Memory Channel to Store Measurement Value
There are 100 directories to store measurement value from M000 to M100 and each can
store one value, therefore 101 in total.
Here, whether measurement value is to store automatically or by pressing
MEMORY
key is
selectable for every measurement.
See p.50 for details on protection of stored data when data storage is to perform on
memory channel with another measurement value.
∗
Setting at the shipment from factory : AUTO NUNM
Operation Procedure
1.
Press
ESC
key when menu or target
value setting menu appears.
Measurement screen
appears on LCD screen.
2.
Make sure that [ ] appears on left
bottom indicating this instrument is
has not been set for SHIFT mode.
If
[SFT]
appears instead, it means that
instrument is to operate in SHIFT mode.
Press
SHIFT
key to cancel SHIFT mode.
3.
Press [MENU] key twice.
Menu 2/4 screen
appears on LCD screen.
4.
Press either
or
key to select
[MEMORY MODE] and then
ENTER
key.
Inversion cursor moves from
[MEMORY MODE]
to right to change
update method of memory channel.
2
3
1, 7
6
4, 5
M
E
M
O
R
Y
M
O
D
E
CH00:
<
DEFAULT
>
Lv 20.80
cd/m
2
x 0.4476
y 0.4477
<
MODE
>
SINGLE
[ ]
=
[ 2° ]
<
ANGL
>
1
°
VIEW
<
SPD
>
AUTO
<
SYNC
>
200.00Hz
<
LENS
>
STANDARD
<
MEM
>
AUTO NUM
<
MEMORY DATA
>
M
OOO
:(NO DATA)
Lv ------
x ------
y ------
<
MENU
>
2/4
OBJECTIVE LENS
STANDARD
DISPLAY DIGITS 4 FIGURES
MEMORY MODE AUTO NUM
DATA PROTECT ON
BREAK : [ESC]
[ ]
=
[ 2° ]
<
MENU
>
2/4
OBJECTIVE LENS STANDARD
DISPLAY DIGITS 4 FIGURES
MEMORY MODE
AUTO NUM
DATA PROTECT ON
BREAK : [ESC]
[ ]
=
[ 2° ]
CS-200_E027-060.indd 50
12.7.17 4:46:24 PM
Summary of Contents for CHROMA METER CS-200
Page 21: ...19 Installing CS 200_E001 026 indd 19 12 7 17 4 43 25 PM...
Page 29: ...27 Setting CS 200_E027 060 indd 27 12 7 17 4 46 00 PM...
Page 57: ...55 Setting BACK LIGHT CS 200_E027 060 indd 55 12 7 17 4 46 28 PM...
Page 62: ...60 CS 200_E027 060 indd 60 12 7 17 4 46 32 PM...
Page 63: ...61 Measurement Preparation CS 200_E061 86 indd 61 12 7 17 4 47 50 PM...
Page 88: ...86 CS 200_E061 86 indd 86 12 7 17 4 48 17 PM...
Page 89: ...87 Measurement CS 200_E087 96 indd 87 12 7 17 4 49 12 PM...
Page 98: ...96 CS 200_E087 96 indd 96 12 7 17 4 49 24 PM...
Page 99: ...97 Communication CS 200_E97 100 indd 97 12 7 17 4 51 08 PM...
Page 102: ...100 CS 200_E97 100 indd 100 12 7 17 4 51 09 PM...
Page 103: ...101 Description CS 200_E101 118 indd 101 12 7 17 4 52 11 PM...
Page 124: ...9222 1892 A1 2005 KONICA MINOLTA INC BIIBPK Printed in Japan...