88
Me
asu
rem
ent
Measurement
1.
Decide whether you use close-up
lens (optional) or not according to
measuring object size and distance.
See the table below for details on measuring distance
and measuring area. If you set close-up lens, lens
type setting is required in this instrument. (see p.40)
2.
Slide measuring angle selector
according to size of object and select
either aperture1˚, 0.2˚, or 0.1˚.
∗
The measuring angle select is covered with holding cap
CS-A24 (see p.8) at shipment from factory.
Pressing with finger cushion, slide
measuring angle selector slowly.
Do not operate measuring angle selector during
measurement. If measuring angle is switched during
measurement, you may fail to perform measurement
or to obtain correct measurement value.
3.
Rotate diopter adjustment ring on
finder for diopter adjustment.
Make sure that aperture (black circle indicating
measurement area) looks clear when
observing object through finder. (See p.17.)
4.
Rotate focus adjustment ring for
objective lens for that purpose.
Make sure that object around aperture looks
clear when observing object through finder.
Only measuring area for measuring object must
be placed in aperture. If extra area, which is
not measuring object, is included in aperture,
you cannot perform correct measurement.
Operation Procedure
1°
0.2°
0.1°
(unit:mm)
Minimum measuring area Maximum measuring area
Minimum measuring
distance
Maximum measuring
distance
Measuring area at 500 mm Measuring area at1000 mm
(Measuring angle) 1
˚
0.2
˚
0.1
˚
1
˚
0.2
˚
0.1
˚
1
˚
0.2
˚
0.1
˚
1
˚
0.2
˚
0.1
˚
1
˚
0.2
˚
0.1
˚
1
˚
0.2
˚
0.1
˚
Without a Close-Up Lens 4.7
1.0
0.5
∞
∞
∞
296
∞
8.5
ø1.7 ø0.9 ø17.7 ø3.6 ø1.8
Close-up Lens No.122
2.2
0.5
0.3
4.6
1.0
0.5
128
240
−
−
−
−
−
−
Close-up Lens No.107
0.8
0.2
0.1
1.1
0.3
0.2
43
52
−
−
−
−
−
−
∗
Measuring distance is the distance from the front edge of the metal lens barrel or close-up lens ring.
Measuring distance and measuring area
Summary of Contents for CHROMA METER CS-200
Page 21: ...19 Installing CS 200_E001 026 indd 19 12 7 17 4 43 25 PM...
Page 29: ...27 Setting CS 200_E027 060 indd 27 12 7 17 4 46 00 PM...
Page 57: ...55 Setting BACK LIGHT CS 200_E027 060 indd 55 12 7 17 4 46 28 PM...
Page 62: ...60 CS 200_E027 060 indd 60 12 7 17 4 46 32 PM...
Page 63: ...61 Measurement Preparation CS 200_E061 86 indd 61 12 7 17 4 47 50 PM...
Page 88: ...86 CS 200_E061 86 indd 86 12 7 17 4 48 17 PM...
Page 89: ...87 Measurement CS 200_E087 96 indd 87 12 7 17 4 49 12 PM...
Page 98: ...96 CS 200_E087 96 indd 96 12 7 17 4 49 24 PM...
Page 99: ...97 Communication CS 200_E97 100 indd 97 12 7 17 4 51 08 PM...
Page 102: ...100 CS 200_E97 100 indd 100 12 7 17 4 51 09 PM...
Page 103: ...101 Description CS 200_E101 118 indd 101 12 7 17 4 52 11 PM...
Page 124: ...9222 1892 A1 2005 KONICA MINOLTA INC BIIBPK Printed in Japan...