Calibration
Calibration Channel
There are 21 calibration channels from CH00 to CH20.
Following settings are available for all channels.
(1) Correction coefficient of user calibration
(2) Target color
(3) CH ID name
These are commonly used among each measurement mode of L
v
xy, L
v
u’v’, L
v
T
∆
uv, XYZ,
and dominant wavelength in one channel.
CH00 is for measurement based upon KONICA MINOLTA’s calibration standard. Its cali-
bration correction coefficient has been set and is unchangeable. Only target color and
CH ID name settings are available.
To switch calibration channel, press either or key in SHIFT mode to switch one ap-
pearing on LCD screen.
CH ID name can be given to each channel through key operation. CH ID name is shown
on LCD screen together with calibration channel.
CS-200_E061-86.indd 62
13.6.28 6:14:14 PM
Summary of Contents for CHROMA METER CS-200
Page 21: ...19 Installing CS 200_E001 026 indd 19 12 7 17 4 43 25 PM...
Page 29: ...27 Setting CS 200_E027 060 indd 27 12 7 17 4 46 00 PM...
Page 57: ...55 Setting BACK LIGHT CS 200_E027 060 indd 55 12 7 17 4 46 28 PM...
Page 62: ...60 CS 200_E027 060 indd 60 12 7 17 4 46 32 PM...
Page 63: ...61 Measurement Preparation CS 200_E061 86 indd 61 12 7 17 4 47 50 PM...
Page 88: ...86 CS 200_E061 86 indd 86 12 7 17 4 48 17 PM...
Page 89: ...87 Measurement CS 200_E087 96 indd 87 12 7 17 4 49 12 PM...
Page 98: ...96 CS 200_E087 96 indd 96 12 7 17 4 49 24 PM...
Page 99: ...97 Communication CS 200_E97 100 indd 97 12 7 17 4 51 08 PM...
Page 102: ...100 CS 200_E97 100 indd 100 12 7 17 4 51 09 PM...
Page 103: ...101 Description CS 200_E101 118 indd 101 12 7 17 4 52 11 PM...
Page 124: ...9222 1892 A1 2005 KONICA MINOLTA INC BIIBPK Printed in Japan...