226
416739/A
e
Select
OK
to save the selected setting and close the dialog box.
Note
If you set
Mode
to Passive, your ES80 will no longer provide any information in the
echogram(s).
2
Enable the
Noise
tooltip.
a
Open the
Display
menu.
b
Select
Display Options
to open the dialog box.
c
Select
Tooltip
to open the page.
d
Select
Noise
to enable the tooltip.
e
Select
OK
to save the selected setting and close the dialog box.
3
For each channel:
a
Place the cursor in the approximate centre of the echogram, and observe the
tooltip information.
b
Read the noise value.
c
Fill in the result table.
Result
Frequency
Noise level
Frequency
Noise level
18 kHz
120 kHz
33 kHz
200 kHz
70 kHz
333 kHz
Requirements
Results
The noise levels are recorded.
Date and signature:
Related topics
Test procedures, page 221
Setting to work, page 153
Simrad ES80 Installation Manual