Maintenance and Troubleshooting

Maintenance

To ensure proper operation, conduct periodic inspection at least 

once a year. 

If the self-diagnostic function displays an alarm that does not 

clear even though the cause of the alarm is eliminated, request 

servicing from Koganei. Contact the Overseas Department for 

more information. 
The addresses and telephone numbers are shown at the back of 

this manual. 

Troubleshooting

● Abnormalities and remedies

Symptom

Remedy

Nothing appears on the 

display. 

 Make sure the supply power voltage and polarity are being 

applied correctly. 

 Make sure that connectors are connected correctly. 

The flow rate display 

does not become zero 

even when valves 

upstream or down-

stream from the flow 

meter is closed and 

there is no flow. (Output 

signal is not 4 mA. )

 Make sure there are no air leaks in the piping. 

 Make sure that wiring is correct. 

 When the flow meter is installed on horizontal piping with the 

display sideways, influence of the installation attitude may cause 

zero point drift.  (Since installation on horizontal piping with the 

display sideways causes the accuracy guarantee flow rate range 

to become 5 to 100% FS, set the low flow cut in this case to 5%. )

 In a location subject to large ambient temperature fluctuation and 

large flow fluid temperature fluctuation, convection within the 

piping may be detected as flow. Perform measurement after 

temperature has stabilized sufficiently. 

Large instrument error 

compared against the 

standard instrument. 

The instrument is 

outside the accuracy 

range. 

Make sure there are no air leaks in the piping.

 Check for foreign matter trapped in the main channel orifice. If 

foreign matter is present, remove it.

 Check for dirt, oil, or other foreign matter in piping and the flow 

meter connection port. If foreign matter is present, Koganei 

servicing and replacement are required. Contact your nearest 

Koganei sales office or the Overseas Department.

 Check to make sure that wiring is correct.

 Check to make sure that there is no large fluctuation in flow rate 

within a few seconds and that the measurement range is not 

exceeded by a large amount. 

● Alarm displays and required actions

 

Alarm

Code

Item

Description

Cause

Remedy

Flow rate over

A flow rate that 

exceeds the value 

set by the maximum 

display value 

parameter setting is 

flowing. Or a back 

flow that exceeds 

the back flow rate 

range is flowing. 

There may be 

a high flow 

rate outside of 

the maximum 

display range 

or back flow. 

Note 1

Check if a high flow rate exists 

and eliminate its cause. If the 

instantaneous flow rate display 

when the alarm occurs shows 

a minus (–) sign, back flow 

occurs. Eliminate the cause of 

the back flow. The alarm will 

cease displaying when the 

actual flow rate becomes within 

the display range flow rate. 

Registration 

data abnormal-

ity 1

Abnormal channel 

ID numbers 

registered. Correct 

flow rate calculation 

cannot be 

performed. 

The registered 

channel ID 

numbers may 

be wrong. 

Use maintenance mode to 

register the correct channel ID 

numbers. 

Note 2

 If changing the 

setting does not clear the 

alarm, turn the instrument off 

and then back on again. If this 

does not clear the alarm, 

request servicing. 

Registration 

data abnormal-

ity 2

Correct calculation 

cannot be 

performed because 

of abnormal data 

registered for the 

flow sensor 

characteristic 

values. 

Registered 

characteristic 

data may be 

corrupted, 

resulting in 

abnormal 

values. 

Turn the instrument off and 

then back on again. If this 

does not clear the alarm, 

request servicing. 

Sensor 

abnormality 1

Flow rate signal 

is outside the 

proper range. 

The signal level 

may be exceedingly 

low due to a short 

circuit caused by 

adherence of 

foreign matter or 

excessive back flow 

that exceeds the 

measurement 

range. 

If excessive back flow is 

the cause, this error will 

clear automatically when 

the flow rate returns within 

the measurement range. If 

it is not due to excessive 

back flow and the alarm 

does not clear after a 

number of hours, request 

servicing. 

Sensor 

abnormality 2 Flow rate signal 

level may not 

match the actual 

flow rate. 

Sensor may be 

malfunctioning or 

output may be 

dropping. Foreign 

matter may be 

adhering to sensor, 

or there may be 

condensation on 

the sensor, etc. 

Pass dry air through the 

system. If the error does 

not clear after a number of 

hours, request servicing. 

Sensor 

abnormality 3

Sensor 

abnormality 4

Heater voltage is 

outside the 

proper range. 

The sensor may be 

damaged, or there 

may be a short 

circuit due to the 

adherence of 

foreign matter, etc. 

If the error does not clear after 

a number of hours, request 

servicing. 

Memory abnormality 

Instrument 

information 

data 

abnormality

Sum mismatch 

when accessing 

EEPROM. 

Data may 

have been 

corrupted by 

noise, etc. 

Turn the instrument off and 

then back on again. If this 

does not clear the alarm, 

request servicing. 

Characteristic 

value information 

data abnormality

Setting 

information 

data 

abnormality

Power was cut 

off during data 

writing, which 

created the 

risk of 

abnormal 

completion. 

Reconfigure the data, and 

then turn the instrument off 

and then back on again. If 

this does not eliminate the 

alarm, request servicing. 

Integrated 

information 

data 

abnormality

Perform an integrated value 

reset, and then turn the 

instrument off and then back 

on again. If this does not 

eliminate the alarm, request 

servicing. 

Note 1:  If parameter setting were used to change the maximum display value, 

this alarm occurs in accordance with the changed setting. 

2:  For information about channel ID numbers, see "Bypass Unit 

Replacement Guidelines" on page 15. 

Summary of Contents for FA Series

Page 1: ...FA Series Flow Meter 注 bタイプの使用サイズは左右 50mm 120mm未満 このデータからの縮率は 50 120 未満 M140787 Ver 3 0 Instruction Manual Ver 3 0 Details ...

Page 2: ...pecifications 16 Order Codes 17 Pressure Loss 18 Dimensions 19 Thank you for purchasing a Koganei FA Series Flow Meter To ensure safe use of this product be sure to read the contents of this manual and gain an understanding of how to properly operate the product before actually trying to use it Keep this manual in a safe place for future reference Contents ...

Page 3: ...ing work etc to mechanisms attached to the product while the product is operating Abnormal operation creates the risk of personal injury WARNING Do not use the product in excess of its specification range Doing so creates the risk of product breakdown loss of function or damage It could also drastically reduce the product s operating life Before supplying air or electricity to the device and befor...

Page 4: ...anei parts or compatible parts recommended parts Always observe the prescribed methods and procedures 2 Never attempt inappropriate disassembly or assembly of the product relating to basic configurations or its performance or functions Koganei shall not be held responsible for any problems that occur as a result of these items not being properly observed General Precautions Mounting location Avoid...

Page 5: ...g piping apply an appropriate amount of sealant Do not apply sealant to the first two threads of the end of the pipe Over application of sealant can cause it to get inside the piping creating the risk of measurement error and damage to the FA Series Flow Meter 7 If different diameter piping a regulator filter valve or other piping equipment is installed upstream from the FA Series Flow Meter be su...

Page 6: ...straight pipe is straight piping that is the same bore diameter as the flow meter The following types of pipes are suitable as straight pipe carbon steel pipes for ordinary piping JIS G3452 carbon steel pipes for pressure service JIS G3454 Schedule 40 or less and stainless steel pipes JIS G3459 Schedule 40 Note 1 For equipment not included in the table below provide 15D or larger straight piping w...

Page 7: ...ctor separately from power lines and high voltage lines use a separate conduit Noise from these lines may cause the equipment to operate erratically 3 When connecting a cable with connector align the core wire and push it in as far as it will go Next tighten the connector nut by hand When tightening the nut be sure to keep it within the prescribed torque 0 4 to 0 6N m range Failure to do so create...

Page 8: ...grated value reaches 0 The integrated value is saved to non volatile memory at 10 minute intervals Note that this means depending on the power off timing the latest maximum 10 minute interval integrated value may not have been recorded Also note that there is no recording for an integrated countdown To reset the integrated value hold down the and keys at the same time for at least five seconds whi...

Page 9: ...s flow rate switch 1 parameter is exceeded Instantaneous flow rate lower limit switch Outputs when the value set with the instantaneous flow rate switch 1 parameter is subceeded Instantaneous flow rate range switch Outputs when instantaneous flow rate is within the range set by instantaneous flow rate switch 1 and instantaneous flow rate switch 2 Integrated flow rate count up switch Outputs when t...

Page 10: ...eys by pressing with a mechanical pencil a screw driver or any other pointed object Doing so can lead to malfunction CAUTION Function settings While the basic screen is displayed hold down the key for about two seconds The three digits on the left will show the function number while the two digits on the right will show the current setting of the function Example of display Function number Current...

Page 11: ...ious unit They are not converted automatically After changing the unit reset integration before using it again The peak value and lower value are reset when the unit is changed Switch output 00 Not used 01 Instantaneous flow rate upper limit 02 Instantaneous flow rate lower limit 03 Instantaneous flow rate range 04 Instantaneous flow rate upper limit inverted 05 Instantaneous flow rate lower limit...

Page 12: ...t cannot be obtained when the setting values is less than 10 of FS Setting zero outputs fixed output during an alarm Reconfigure this setting after changing the display unit setting with function selection Analog output 4mAflow rate specificationNote 1 Flowratevalue equivalentto0to400 FScanbeset Note5 0 Settings are values with decimal points removed Decimal points will be added with model setting...

Page 13: ... flow rate screen is displayed hold down the key for about two seconds This will change to the instantaneous flow rate lower value screen which alternate between and the lower value To exit the lower value screen press the key This will return to the instantaneous flow rate screen or integrated flow rate screen To clear the current instantaneous flow rate lower value display the lower value and th...

Page 14: ...ure compensation Channel ID number 0 Adjustment value Configured when using the replacement bypass unit Settings register main channel characteristic values Not displayed for FAS 002 FAS 005 Do not change except when replacing bypass unit Cannot be changed without inputting a password for the reservation code Channel ID number 1 Adjustment value Channel ID number 2 Adjustment value Model FAS 030 2...

Page 15: ...ck flow The alarm will cease displaying when the actual flow rate becomes within the display range flow rate Registration data abnormal ity 1 Abnormal channel ID numbers registered Correct flow rate calculation cannot be performed The registered channel ID numbers may be wrong Use maintenance mode to register the correct channel ID numbers Note 2 If changing the setting does not clear the alarm tu...

Page 16: ...mber label of the main channel ー ー Body Number Channel ID number 0 Channel ID number 1 Channel ID number 2 Model numbers are 2 for the FAS 030 3 for the FAS 060 and 4 for the FAS 120 Be sure to double check the channel ID numbers and the model number after you input them 4 Check to make sure that the instantaneous flow rate display shows zero 5 Apply pressure to the flow meter and inside the pipin...

Page 17: ...01 MPa 1 digit 5 to 100 of FS flow 1 current output 4 to 20 mA Output range 3 2 to 20 8 mA Fixed output during alarm up 21 6 0 4 mA initial default modifiable by setting Fixed output during alarm down 0 0 0 4 mA fixed Allowable load resistance 300 Ωmax Maximum output current 24 mA max Within 1 5 s For 0 100 FS step input time to reach 95 of maximum value Calculation cycle 50 ms Rc 1 4 Rc 1 2 Rc1 R...

Page 18: ...channel material Bypass unit Gasket Bypass unit case material Load mass kg Semiconductor non volatile memory EEPROM Recorded data Function settings parameters integrated values etc 50 MΩ min Between connector pin and main channel or mounting bolts by 500VDC megger Option with connector oil resistant flame retardant cable UL2464 EN standard See Installation attitudes on page 4 IP65 JISC0920 and IEC...

Page 19: ...5 0 1 MPa Primary side pressure 0 7 MPa 0 3 MPa 0 5 MPa 0 1000 6000 5000 4000 0 2000 3000 Flow rate R min normal 0 1 MPa Primary side pressure 0 7 MPa 0 3 MPa 0 5 MPa 0 80 20 40 60 Pressure loss kPa Pressure loss kPa FAS 060 20 5 10 15 0 1 MPa Primary side pressure 0 7 MPa 0 3 MPa 0 5 MPa 0 1000 6000 5000 4000 0 2000 3000 Flow rate R min normal 0 1 MPa Primary side pressure 0 7 MPa 0 3 MPa 0 5 MPa...

Page 20: ...threaded mounting hole 2 M4 x 0 7 depth 7 87 8 90 78 75 63 4 5 5 5 Mounting bracket B FAS 030 FAS 002 2 Rc1 4 FAS 005 2 Rc1 2 threaded mounting hole 2 M4 x 0 7 depth 7 46 5 15 34 75 47 13 5 24 48 27 65 8 1 5 73 8 23 5 IN OUT φ6 90 78 75 63 4 5 5 5 Mounting bracket B M12 connector Cable with connector option 2NL Cable length 2000 mm 5NL Cable length 5000 mm FAS 002 FAS 005 ...

Page 21: ...th 2000 mm 5NL Cable length 5000 mm φ6 M12 connector threaded mounting hole 2 M5 x 0 8 depth 7 IN OUT FAS 060 2 Rc2 34 17 61 30 5 110 106 8 75 37 5 75 87 5 37 5 Cable with connector option 2NL Cable length 2000 mm 5NL Cable length 5000 mm φ6 M12 connector threaded mounting hole 2 M5 x 0 8 depth 7 OUT IN FAS 120 ...

Page 22: ...φ4 5 Cross head pan screw M4 0 7 length 6 2 pcs supplied 48 78 90 Additional Parts Dimensions mm φ6 φ14 40 40 FAK 2L 2000 FAK 5L 5000 Bypass unit FAB Mounting bracket B FAS Cable with connector option FAK 2L Cable length 2000 mm Yamatake Corporation FAK 5L Cable length 5000 mm ...

Page 23: ...tact the Overseas Department below Inquiries Koganei Corporation Overseas Department 3 11 28 Midori cho Koganei City Tokyo 184 8533 TEL 042 383 7172 FAX 042 383 7206 Revision History Ver 2 0 P17 The order codes of Lenght of cable with connector will be changed as bellow 2L 2NL 5L 5NL The order code of cable with connector in additional parts will be changed as below PA5 41S 2MK E FAK 2L PA5 41S 5M...

Page 24: ...hina Tel 021 6145 7313 Fax 021 6145 7323 KOGANEI PORNCHAI CO LTD 89 174 Moo 3 Vibhavadee Rangsit Road Talad Bangkhen Laksi Bangkok 10210 Thailand Tel 02 551 4025 Fax 02 551 4015 KOGANEI KOREA CO LTD Room 301 Donghwa Bld 25 5 Yoido Yongdeunpo Gu Seoul Korea Tel 02 786 0413 Fax 02 786 0415 For other information and more detailed specifications see the product catalog For inquiries about the product ...

Reviews: