4. The set indication 

cannot be obtained. 

a. The pressure on the inlet side of the 

mass flow controller is too high or the 

controller is faulty. Or, a pressure 

difference between the inlet side and 

outlet side is not adequate. 

1. Adjust the pressure to the adequate level as shown 

in the brochure or set an adequate differential 

pressure. 

b. The pressure loss of piping is large. Or 

the pressure loss has increased as the 

filter, check valve, etc. has become nearly 

clogged. 

2. Set a pressure gauge immediately before and after 

the mass flow controller and check to see if the 

pressure difference is adequate. Make adjustment to 

get an adequate differential pressure. 

c. The orifice in the control valve is almost 

clogged. 

3. If the problem is solved by increasing the pressure, 

it is most likely that the orifice is clogged. To replace 

the orifice, please send it to KOFLOC. 

5. The flow rate is clearly 

less that the indicated flow 

rate. 

a. The piping line is leaking. 

1. Using the evacuation method, leak detect fluid, etc., 

investigate causes of leak such as failure to tighten 

piping line nuts, insufficient tightening of such nuts or 

others and retighten the nuts of leaking places or take 

other corrective actions to stop leak. 

b. The mass flow main unit is leaking. 

1. Same as above. When checking leak using leak 

detect fluid, apply such fluid only to the joints before 

and after the unit, but do not apply it to the inside of 

the main unit cover. After checking, wipe off leak 

detect fluid completely so that it will not remain at all. 

c. The bypass is clogged. 

1. To replace the bypass part, please send it to 

KOFLOC. 

6. The flow rate is clearly 

larger than the indicated 

flow rate. 

a. The sensor tube is nearly clogged. 

1. To replace the sensor tube, please send it to 

KOFLOC. 

7. Even if the command is 

set to zero, gas keeps 

flowing. 

a. The equipment is not in the analog 

setting mode. Or, the external input set 

voltage is not zero. Or, the printed board is 

faulty. Or, the wiring of the command 

soldered part is broken. 

1. Change the equipment to the analog setting mode 

and check the operation. If the operation will not 

become normal, check the external input set voltage. 

2. Input the valve forced close signal GND to Pin No. 8 

and see if gas stops. If gas does not stop, the valve needs 

to be readjusted. Please send it to KOFLOC. 

If gas stops, a possible cause is a faulty electronic 

circuit. Please send it to KOFLOC. 

b. Internal leak due to adhesion of dust, 

etc. to the sealing area of the control valve. 

1. To overhaul and clean the control valve, please send 

it to KOFLOC. 

c. The zero point voltage has moved to the 

negative side. 

1. The zero point voltage is abnormal. See Symptom 2. 

The output will not become zero. 

Summary of Contents for 8700 MC

Page 1: ...MJ400306A1 1 Thermal Mass Flow Meter Controller MODEL 8700 SERIES INSTRUCTION MANUAL...

Page 2: ...improvement Prior to shipment every care has been taken in preparing this manual not to mention the product itself but if you notice any imperfections errors or omission please contact KOFLOC Prior to...

Page 3: ...ion Prior to Use 14 9 Nomenclature and Functions of the Indicator 15 10 Nomenclature and Functions of the Main Unit 16 11 Basic Operation of the Indicator 17 12 Changing Over the Operation Modes 20 13...

Page 4: ...ure that the proof pressure on the input equipment side is sufficient The output may possibly be in a range of 1 to 7 V in the case of voltage and 0 to 26 mA in the case of current 2 Items to Check ab...

Page 5: ...s may occur When it is expected that the product will be kept in storage for a long time take the following precautions Store the product in the package in which it was received from KOFLOC wherever p...

Page 6: ...large flow rate realized by a small body by employing a high lift actuator Controller type An indicator is installed and can be operated with a single 24 VDC power supply only Both of digital control...

Page 7: ...S 5SLM 150 300kPa F S 50SLM 200 300kPa F S 50SLM Pressure characteristic 0 3 F S 100 kPa Temperature Working temperature 5 45 Temperature characteristic 0 1 F S 0 2 F S Humidity Allowable operating h...

Page 8: ...US316 SUS316L SUS430 Fluor rubber PTFE Chloroprene rubber Optional Joint Rc1 4 Rc3 8 Optional 1 4SWL 1 8SWL 1 4VCR 3 8SWL Weight Integrated type approx 1200 g separate type approx 1400 g Integrated ty...

Page 9: ...d flow rate for one minute maximum from the storage to the power failure is lost 3 Upper lower limit flow rate alarm allowable range OK judgment function To use this function refer to 13 Applied Opera...

Page 10: ...in a place where ambient temperature is 15 to 35 The use of the equipment in a condition deviating from the conditions described in the specifications is a cause of failure When installing the mass f...

Page 11: ...to see if the connections of gas piping are not leaking If such a check is omitted and unsafe gas is used serious accidents may occur Give careful attention to piping after the unit also to ensure tha...

Page 12: ...6 Flow rate setting input 1 0 5 V 4 20 mA 7 Flow rate output 20 5 V 4 20 mA 8 Operation mode input 5V FULL OPEN S GND CLOSE OPEN CONTROL 9 5V output 5V Valve motion mode input only 10 ALARM output NPN...

Page 13: ...13 Do not connect different power supplies to the power input terminals of the above connector and the pin jack power terminals simultaneously Such a practice may cause a fire or damage to the equipm...

Page 14: ...ero adjust function Pressing the ZERO switch on the side of the main unit corrects deviation of the zero point This step is not needed when the auto zero function is active 3 Introducing gas After con...

Page 15: ...ey Digit Used to select a digit of the indicated value to change when setting a flow rate and various functions RUN key Used to change over the operation mode between fully closed control fully opened...

Page 16: ...hts when the event output 2 is on 7 segment LED 4 digits Indicates various flow rates and function set modes 10 Nomenclature and Functions of the Main Unit Side view of the main unit Never peel off th...

Page 17: ...w indication Not shown when the upper SV lamp on DISP key or mid upper values of the integrated Upper of integ flow indication flow rate value are zero SV lamp on DISP key Instantaneous flow rate indi...

Page 18: ...usly every 0 5 second Determination of a numeric value After getting a desired numeric value by the above operation press the ENT key The flashing will stop and the set value will be indicated When th...

Page 19: ...vided to 4 digits each and indicated three times Resetting the integrated flow rate When an integrated value is being indicated TF lamp lit pressing and holding the ENT key for 3 seconds or longer res...

Page 20: ...ode to the fully opened mode press and hold the RUN key for 2 seconds or longer Control mode RUN key less than RUN key 2 sec less than 2 sec Fully closed mode RUN key 2 sec or longer RUN key less than...

Page 21: ...ng in D 01 Flow rate setting method selection in 14 List of Function Setting Modes six SP values SP0 5 can be changed over by event inputs as shown below In the case of no event input or multiple even...

Page 22: ...blue line Upper limit alarm When the PV value exceeded the SV value AL Hi the flow rate alarm is turned on after the passage of AL dt Lower limit alarm When the PV value dropped below the SV value AL...

Page 23: ...AL Hi Flow rate alarm upper limit value D 15 AL Lo Flow rate alarm lower limit value D 16 AL dt Flow rate alarm delay time D 17 PV Instantaneous flow rate Upper limit alarm When the PV value exceeded...

Page 24: ...r limit alarm value the OK LED lights Refer to D 18 to D 21 in 14 List of Function Setting Modes 6 Using the integration setting function Integration event occurrence When an integration event flow ra...

Page 25: ...other than SP 02 Key lock of all settings D 07 RUN key action selection 01 00 RUN key not used 01 RUN key used D 08 Auto zero function 00 00 0FF 01 0N 01 The zero point is automatically adjusted when...

Page 26: ...alarm upper limit value F S 10 0 Settable range 0 5 100 0 For Controller 100 0 Settable range 0 0 100 0 For Meter D 16 Flow rate alarm lower limit value F S 10 0 Settable range 0 5 100 0 For Controlle...

Page 27: ...6 times 05 Moving average of sampling 32 times D 28 Indication update cycle indicator 02 00 No update cycle 01 50 msec cycle 02 100 msec cycle 03 200 msec cycle 04 500 msec cycle 05 1000 msec cycle D...

Page 28: ...le range 0 to full scale D 37 SP2 Expanded set flowrate 0000 Settable range 0 to full scale D 38 SP3 Expanded set flowrate 0000 Settable range 0 to full scale D 39 SP4 Expanded set flowrate 0000 Setta...

Page 29: ...top the supply of gas set the operation mode to the fully closed mode or reset the set flow rate to zero This will prevent counterflow AL 4 Upper limit alarm Insufficient flow rate alarm delay time un...

Page 30: ...leaned or replaced Please send it to KOFLOC h The sensor tube is clogged 1 If the tube is clogged gas keeps flowing Check it The sensor needs to be replaced Please send it to KOFLOC 2 The output will...

Page 31: ...in unit is leaking 1 Same as above When checking leak using leak detect fluid apply such fluid only to the joints before and after the unit but do not apply it to the inside of the main unit cover Aft...

Page 32: ...it will be charge free repaired in our factory But if a malfunction of the machine occurs due to the following reasons even within the warrant period it becomes the outside for a warranty a Malfuncti...

Page 33: ...MJ400306A1 33 19 External View 1 Integrated indicator type...

Page 34: ...MJ400306A1 34...

Page 35: ...MJ400306A1 35 2 Separate indicator type Main unit...

Page 36: ...MJ400306A1 36...

Page 37: ...MJ400306A1 37 Indicator URL http www kofloc co jp...

Reviews: