Basic Operation of the Indicator
(1)
Changing over the indication of the instantaneous flow rate, set flow rate and integrated
flow rate
Power on
↓
▼ key+ENT key held pressed
Status of lamps on the indicator
①
Instant. flow indication
⑧
Function set mode
([PV] and [SV] lamps on)
↓DISP key
DISP key
②
SP No. set
[SV] lamp on
↓DISP key
③
Set flow indication
[SV] lamp on
↓DISP key
④
Lower of integ. flow indication
[SV] lamp on
↓DISP key
⑤
Mid of integ. flow indication
Not shown when the upper
[SV] lamp on
↓DISP key
or mid & upper values of the integrated
⑥
Upper of integ. flow indication
flow rate value are zero.
[SV] lamp on
DISP key
①
Instantaneous flow rate indication
The present flow rate is indicated.
The unit of indication is based on the ordered scaling (ex. 1.000 in the case of 1 SLM).
When the operation mode is changed, the operation mode is indicated for a certain
time.
②
SP No. set(Controller type only)
This equipment can hold six SPs (set points) maximum.
※
When the SP No. is being shown, if no key is operated for 10 seconds, the indication
will automatically change to the instantaneous flow rate.