Koehler K2269X Operation And Instruction Manual Download Page 10

K2269X Low Temperature Pressure Viscometer 

Operation and Instruction Manual 

 

K2269X-Manual

    

 

-

6-

 

build-up  of  the  system  and  rupturing  of  the 
over-pressure emergency valve. Always make 
sure  opposing  valve  is  in  the  open  position 
prior to closing a valve.  
 

9.  Refrigerated  Chamber  Lid. 

Keep  lid  in 

closed  position  to  maintain  temperature 
uniformity 

of  the  refrigerated  chamber. 

Remove  lid  when  inserting  and  removing  the 
grease cylinder.  

 

10.  Right-Hand Needle Valve Knob: 

Part of dual 

valve 

hydraulic 

system 

designed 

for 

continuous  flow  of  hydraulic  fluid.  Turning 
Right-Hand  knob  counter-clockwise  opens 
valve and allows for fluid flow to pressurize the 
grease  cylinder.  Closing  Right-Hand  valve  by 
turning  knob  clockwise  closes  the  path  to  the 
grease  cylinder  and  forces  the  fluid  to  flow 
through the continuous system loop.  

 

NOTE:

 The two valve system (Left and Right-

hand  Knobs)  must  be  used  in  opposition. 
When  one  valve  is  open,  the  other  must  be 
closed, and vice-versa.  

 

WARNING:

  Never  close  both  valves  at 

the  same  time  as  this  will  result  in  pressure 
build-up  of  the  system  and  rupturing  of  the 
over-pressure emergency valve. Always make 
sure  opposing  valve  is  in  the  open  position 
prior to closing a valve.  

 

11.  Hydraulic  System  Connection  Port: 

For 

connection to hydraulic system. Quick connect 
coupling allows the user to easily connect the 
grease  cylinder  to  the  hydraulic  system.  This 
port is also used as the main connection point 
for the hydraulic system calibration by utilizing 
the  Calibration  Pipe  Assembly,  See  Figure  1, 
Item G.    

 
 
 
 
 
 
 
 
 
 
 
 
 
 

4.3  Control Box Descriptions 

 

 

 

Figure 4. Instrument Descriptions_Control Box 

 

12.  Compressor  Switch. 

Controls  power  to 

refrigeration system. Turning on the switch will 
also  power  on  the  lamp  in  the  receiver 
chamber as well as the Chamber Temperature 
Controller 

(16)

  and  Sample  Temperature 

Display 

(17)

.  

 
NOTE: 

When  power  to  the  compressor  is 

turned  off,  the  user  must  wait  for  a  period  of 
30 minutes prior to turning the power back on. 
This  is  to  prevent  any  damage  to  the 
compressor. 

 
13.  Drive  Motor  /  Circulating  Pump  Switch. 

Turns  on  both  the  drive  motor  which  powers 
the    hydraulic  /  gear  pump  as  well  as  the 
circulating  pump  which  pushes  refrigerant  to 
the hydraulic / gear pump.  

 

NOTE:

  It  is  not  necessary  to  switch  on  the 

Drive  Motor  /  Circulating  Pump  until  the 
instrument  has  reached  the  proper  test 
temperature.  

 

14.  Reset  Button. 

The  gear  box  cover  is 

equipped  with  a  safety  switch  that  shuts  off 
the  Drive  Motor  /  Circulating  Pump  when 
opened. If the gear box cover is opened for an 
extended  period  of  time  the  Drive  Motor  / 
Circulating  Pump  may  not  automatically  be 
switched back on when the cover is closed. In 
this  case,  pressing  the  Reset  Button  will  turn 
the Drive Motor / Circulating Pump back on.  

 

15.  Gear Switch.

 Allows for changing drive motor 

speed  to  simulate  flow  rates  attained  by  40-
tooth  gear  and  64-tooth  gear  as  per  ASTM 
D1092. 

 

 

Summary of Contents for K2269X

Page 1: ...enue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis...

Page 2: ......

Page 3: ...tured in conformance with the applicable standards set forth in this certification Specifications ASTM D1092 This unit is tested before it leaves the factory to ensure total functionality and complian...

Page 4: ......

Page 5: ...tion 5 4 1 Instrument Descriptions 5 4 2 Top Assembly Description 5 4 3 Control Box Descriptions 6 4 4 Gear Box and Motor Descriptions 7 4 5 Temperature Controller Descriptions 7 5 Operation 9 5 1 Ref...

Page 6: ...pport of your laboratory testing needs Our products are backed by our staff of technically knowledgeable trained specialists who are experienced in both petroleum products testing and instrument servi...

Page 7: ...t may exhibit potential hazards Any user must be familiarized with the possible dangers before use We also recommend consulting the Material Data and Safety Sheet MSDS on each chemical reagent for add...

Page 8: ...e unit and all accessories for damage If any damage is found keep all packing materials and immediately report the damage to the carrier We will assist you with your claim if requested When submitting...

Page 9: ...ar pump gears drive motor required to power the hydraulic system Refer to section 4 4 for detailed descriptions on this section of the apparatus 4 2 Top Assembly Description Figure 3 Instrument Descri...

Page 10: ...nect the grease cylinder to the hydraulic system This port is also used as the main connection point for the hydraulic system calibration by utilizing the Calibration Pipe Assembly See Figure 1 Item G...

Page 11: ...n See Figure 4 item 13 Variable Speed Motor is coupled to the speed reducer and is capable of operating at two speeds required to simulate flow rates attained by 40 tooth gear and 64 tooth gear as per...

Page 12: ...a certified calibrated measuring device to acquire the temperature Calculate the difference between the measuring device and the Process value displayed on the controller b Press the program key two t...

Page 13: ...ion instructions for the temperature controller NOTE Allow for the test chamber to cool and come to temperature equilibrium prior to turning on Drive Motor Recirculating Pump See Figure 4 Item 13 It m...

Page 14: ...imed 11 Once a steady flow or constant drip rate is observed the hydraulic system is primed and free of air bubbles 12 Using the relief valve at the right hand side of the calibration piping assembly...

Page 15: ...esting purposes If not attached ferrule will be in thermocouple packaging 3 Remove bottom cap from Grease Cylinder F 4 Invert grease cylinder and place Grease Cylinder Piston C inside Grease Cylinder...

Page 16: ...cylinder loading procedure is now complete Before attaching a capillary changing the flow of the hydraulic fluid and starting a test the sample grease must attain temperature equilibrium with the test...

Page 17: ...may result in personal injury or death If you have any questions about maintaining your equipment then please do not hesitate to contact the Koehler technical service department 6 1 Routine Maintenan...

Page 18: ...104 002 Fuse 25A Slo Blo 5X20 M M 220 240 013 Cartridge Heater 500W 240V 265 500 001 RTD Temperature Probe x 12 275 103 044 Temperature Controller 100 240V 288 230 018 Motor 208 230V 3 Phase 288 230 0...

Page 19: ...K2269X Low Temperature Pressure Viscometer Operation and Instruction Manual K2269X Manual 15 7 Wiring Diagrams 7 1 220 240V 50Hz Unit Wiring K22695...

Page 20: ...e the return of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the re...

Page 21: ...K2269X Low Temperature Pressure Viscometer Operation and Instruction Manual K2269X Manual 17 Notes...

Page 22: ...K2269X Low Temperature Pressure Viscometer Operation and Instruction Manual K2269X Manual 18 Notes...

Reviews: