9.8 Informational Parameters (Read Out Parameters)
P60:
Overall operating hours of the unit (h)
Indication varies according to the elapsed time:
Operating hours
Indication form
0 to 999.9h
xxx,x
1000 to 9999h
xxxx
Over 9999h
X,xx: e meaning x,xx 10
e
P61:
Time elapsed after last switch-on (h)
P62:
Operating hours of the relay (h)
P63:
Number of switching cycles of the relay (h)
Indications are the same as in
P60
.
P64:
Actual temperature of the transducer (°C/°F)
P65:
Maximum temperature of the transducer (°C/°F)
P66:
Minimum temperature of the transducer (°C/°F)
In case of a breaking in the temperature circuit „tErr” will be displayed. The transmitter will perform temperature
correction corresponding to 20ºC.
.
P70:
Number of Echoes / Echo Map
NUS-4 is monitoring the echo
conditions.
Entering this parameter will save
the actual echo map.
Number, distance and amplitude
of these echoes can be read-out
one by one.
yy
: xxxx
yy
: xxxx
70
: _ _ XX
yy
: xxxx
yy
: xxxx
Number of
yy
Serial number
of the echo
xx
value of
DIST or AMPL
echos
(DISTANCE)
(DISTANCE)
(AMPLITUDE)
(AMPLITUDE)
P71:
Distance of the of Measuring Window
P72:
Amplitude of the Echo in the Measuring
P73:
Echo Position (time) :(ms)
P74:
Signal To Noise Ratio
Ratio
Measurement conditions
Over 70
Excellent
Between 70 and 30
Good
Under 30
Unreliable
P75:
Blocking Distance
The actual close-end blocking distance will be displayed (provided automatic blocking was selected in
P05)
.