Kobold HND-P210 Operating Instructions Manual Download Page 7

 

 

HND-P210 

HND-P210 K08/0422 

 

page 7 

Max Memory:  

Pressing ´max´ (key 2) shows the maximum of the measured 
values.  Pressing  it  again  hides  them.  To  clear  the  max 
memory press key ´max´ for >2 seconds.

 

Min Memory:  

Pressing ´min´ (key 5) shows the minimum of the measured 
values.  Pressing  it  again  hides  them.  To  clear  the  min 
memory press key ´min´ for >2 seconds.

 

Hold Function:  

By pressing ´Store/Quit´ (key 6) the last measuring value will 
be  held  in  the  secondary  display.  Pressing  it  again  hides  it. 
(only when logger = ‚off‘).

 

Zero-Point  Adjustment: 

If  there  is  no  pressure  or  zero-pressure  (absolute) 

applied to the pressure ports the device will display 0. If there 
is  a  permanent  deviation  (and  device  is  operated  under 
steady  conditions),  a  permanent  zero  point  adjustment  can 
be carried out. 

 

To  carry  out  the  adjustment  press  button  3  for  approx.  5 
seconds. (Auto Null will be displayed shortly). The adjustment 
is  done  via  the  OFFSET-value  of  the  sensor  (referring 
configuration menu). 

Note: A zero-point adjustment can only be carried out if the 
difference between the value on display is less than 500 
digits!  

To recall the manufacturer`s calibration press button 3 for approx. 
15 seconds. 

Note: If a zero-point adjustment was carried out, this will be 
signalled by the short displaying of „Corr“ when switching on 
the device. 

 
 

8.4  Configuration 

To change device settings, press 

Menu 

(key 4) for 2 seconds. This will call the 

configuration menu (main display: „SEt“).

 

Pressing key 

Menu 

changes between the menus, pressing 

(key 3) jumps to the 

referring parameters, which can be selected with key 

(key 3).

 

The parameters can be changed with 

 (key 2) or 

 (key 5). 

Pressing 

Menu 

again jumps back to the main configuration menu and saves the 

settings. 

Quit 

(key  6)  finishes  the  configuration  and  returns  to  standard  measuring 

operation. 

Summary of Contents for HND-P210

Page 1: ...Operating Instructions for Hand held Pressure Measuring Devices for External Pressure Sensors Model HND P210...

Page 2: ...r Level Display Unit m 11 8 6 Operation Of Logger 11 8 7 The Serial Interface 14 8 8 Pressure Connection to The Sensors 15 8 9 Error and System Messages 16 8 10 Calibration Services 16 9 Maintenance 1...

Page 3: ...iced by persons familiar with these operating instructions and in accordance with local regulations applying to Health Safety and prevention of accidents When used in machines the measuring unit shoul...

Page 4: ...eration With Power Supply Warning When using a power supply please note that operating voltage has to be 10 5 to 12 VDC Do not apply overvoltage Cheap 12 V power supplies often have excessive no load...

Page 5: ...e the device temperature has adjusted to the ambient temperature before trying a new start up 4 If device is to be connected to other devices e g via serial interface the circuitry has to be designed...

Page 6: ...Furthermore the kind of measuring is displayed Slo Fast P det p r t 2 1 and Corr if a Sensor with activated offset or scale correction is connected After changing the battery the clock setting menu is...

Page 7: ...tment press button 3 for approx 5 seconds Auto Null will be displayed shortly The adjustment is done via the OFFSET value of the sensor referring configuration menu Note A zero point adjustment can on...

Page 8: ...mbar Enter desired value at which the analogue output potential should be 1 V if Out dAC Set CAL Set Calibration Adjustment of Sensor OFFS Sensordep e g 5 00 5 00 mbar The offset of sensor will be dis...

Page 9: ...Application with logger function Measuring of short pressure peaks or fast changing pressures with a resolution of 1ms The cyclic logger function records the arithmetic mean value the highest and the...

Page 10: ...rface converter To get access to each device the base addresses of the devices have to be different For example choose 01 for the first 11 for the second device and so on See also chapter 8 7 The Seri...

Page 11: ...tore 2 seconds The logger records 3 measurement results each time current or mean value depending on logger setting see below min peak and max peak of sensor Min and max peak are the minimum resp the...

Page 12: ...etween the different values referring to the measurement also is done by pressing Changing the measurement is done by pressing the keys or 8 6 2 Func CYCL Automatic Recording With Selectable Logger Cy...

Page 13: ...ording Do not stop the recording The selection can be made by key 2 and key 5 Quit key 6 enters the choice Note If you try to switch off the instrument in the cyclic recording operation You will be as...

Page 14: ...me Conf P oFF x 12 Read ID number x 223 Set power off time Conf P oFF x 22 Read min alarm rail AL AL Lo x x 224 Logger Read data of CYCL Logger x 23 Read max alarm rail AL AL Hi x 225 Logger Read cycl...

Page 15: ...e measuring range by re plugging the tube to pressure port A Please note that all values are displayed as positive values No minus sign will be shown Example for HND PS02 For tube connection B the mea...

Page 16: ...Sensor defective Return to manufacturer for repair Err 2 Measured value below allowable range Check pressure not within sensor range measuring value to low Sensor defective Return to manufacturer for...

Page 17: ...ensor Accuracy typ 0 1 F S at nominal temperature FASt and P dET 0 5 FS Measuring rate slow 4 meas sec ConF Rate slow fast 1000 meas sec ConF Rate FASt and P dEt Nominal temperature 25 C Sensor All se...

Page 18: ...consumption Slow measuring rate 1 6mA Fast measuring rate 7 0mA Low Power Logger 0 3mA for cycle time 30s without interface communication active and no alarm horn sounding up to 0 8mA at cycle time 1s...

Page 19: ...HND P210 HND P210 K08 0422 page 19 11 Order Codes Order no Housing design HND P210 1 x pressure sensor input with additional functions...

Page 20: ...S06 0 2000 mbar abs HND PS07 0 7 00 bar abs 10 mbar max 10 bar abs HND PS08 0 400 0 mbar rel 0 1 mbar max 2 bar rel 0 70 C External threads G Edelstahl HND PS19 0 1000 mbar rel 1 mbar max 5 bar rel HN...

Page 21: ...olated HND Z033 Adapter RS232 converter on USB interface HND Z034 Windows software for setting and data read and print out of instruments of the HND series with logger function BUS S20M Software for r...

Page 22: ...ds noted below EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements Part 1 General requirements EN IEC 63000 2018 Technical documentation for the assessment...

Reviews: