Kobold HND-P210 Operating Instructions Manual Download Page 14

HND-P210 

page 14 

 

HND-P210 K08/0422 

8.7  The Serial Interface 

By  means  of  the  serial  interface  and  a  suitable  electrically  isolated  interface 
adapter  (HND-Z031/-032)  the  device  can  be  connected  to  a  computer  for  data 
transfer.  To  avoid  transmission  errors,  there  are  several  security  checks 
implemented e.g. CRC. 

 

The following standard software packages are available: 



HND-Z034

:  Operation  and  read  out  of  logger  function,  data  display  in 

diagrams and tables 



BUS-SW9M

:  9-channel software to display the measuring values

 

 

The device has 3 channels: 

1:  current measuring value (base address) 
2:  min peak (see chapter 

8.6 Operation Of Logger

3:  max peak ( see chapter 

8.6 Operation Of Logger

)

 

Note: The measuring-/ alarm- and display range values read back from 
the interface are always in the selected measurement unit (mbar, bar...)!

 

Supported functions: 

 

Channel 

Code 

Name/Function 

Channel 

Code 

Name/Function 

2, 3 

 

 

1

2, 3

 x 

Read measurement value

x

208

Read # of channels 

 x 

Read system state 

x

222

Read power off time (Conf-P.oFF)

   

12 

Read ID number 

 

223 

Set power off time (Conf-P.oFF) 

   

22 

Read min alarm rail (AL. - AL.Lo) 

224 

Logger: Read data of CYCL- Logger 

   

23 

Read max alarm rail (AL. - AL.Hi) 

 

225 

Logger: Read cycle time (LoGG - 
CYCL)

 

32 

Read configuration flag 

BitAlarmOn:1; BitAlarmSound:3; 
BitLoggerOn:50; BitCyclicLogger:51; 
BitLowPowerLogger:52

 

 

226 

Logger: set cycle time (LoGG - 

CYCL) 

   

160 

Set configuration flag (refer to 32)

x

227

Logger: start recording 

176 

Read min measuring range 

 

228 

Logger: Read # of recordings made 

177 

Read max measuring range 

 

229 

Logger: Read state 

178 

Read measuring range – measuring 
unit 

 

231 

Logger: Read stop time 

179 

Read measuring range – decimal 
point 

 

233 

Read real time clock (CLOC) 

180 

Read kind of measuring of sensor

x

234

Set real time clock (CLOC)

 

 

 

 

x

236

Read logger memory size 

199 

Read kind of measuring of display 

 

237 

Read logger filecount 

200 

Read min display range 

 

238 

Read logger filepointer 

201 

Read max display range 

x

239

Read logger file info 

202 

Read display range - unit 

x

240

Reset

204 

Read display range – decimal point

x

254

Program version 

 

 

 

 

 

260 

Logger: read data of STOR Logger 

 

 

 

 

263 

Read logger channel info 

 

 

Summary of Contents for HND-P210

Page 1: ...Operating Instructions for Hand held Pressure Measuring Devices for External Pressure Sensors Model HND P210...

Page 2: ...r Level Display Unit m 11 8 6 Operation Of Logger 11 8 7 The Serial Interface 14 8 8 Pressure Connection to The Sensors 15 8 9 Error and System Messages 16 8 10 Calibration Services 16 9 Maintenance 1...

Page 3: ...iced by persons familiar with these operating instructions and in accordance with local regulations applying to Health Safety and prevention of accidents When used in machines the measuring unit shoul...

Page 4: ...eration With Power Supply Warning When using a power supply please note that operating voltage has to be 10 5 to 12 VDC Do not apply overvoltage Cheap 12 V power supplies often have excessive no load...

Page 5: ...e the device temperature has adjusted to the ambient temperature before trying a new start up 4 If device is to be connected to other devices e g via serial interface the circuitry has to be designed...

Page 6: ...Furthermore the kind of measuring is displayed Slo Fast P det p r t 2 1 and Corr if a Sensor with activated offset or scale correction is connected After changing the battery the clock setting menu is...

Page 7: ...tment press button 3 for approx 5 seconds Auto Null will be displayed shortly The adjustment is done via the OFFSET value of the sensor referring configuration menu Note A zero point adjustment can on...

Page 8: ...mbar Enter desired value at which the analogue output potential should be 1 V if Out dAC Set CAL Set Calibration Adjustment of Sensor OFFS Sensordep e g 5 00 5 00 mbar The offset of sensor will be dis...

Page 9: ...Application with logger function Measuring of short pressure peaks or fast changing pressures with a resolution of 1ms The cyclic logger function records the arithmetic mean value the highest and the...

Page 10: ...rface converter To get access to each device the base addresses of the devices have to be different For example choose 01 for the first 11 for the second device and so on See also chapter 8 7 The Seri...

Page 11: ...tore 2 seconds The logger records 3 measurement results each time current or mean value depending on logger setting see below min peak and max peak of sensor Min and max peak are the minimum resp the...

Page 12: ...etween the different values referring to the measurement also is done by pressing Changing the measurement is done by pressing the keys or 8 6 2 Func CYCL Automatic Recording With Selectable Logger Cy...

Page 13: ...ording Do not stop the recording The selection can be made by key 2 and key 5 Quit key 6 enters the choice Note If you try to switch off the instrument in the cyclic recording operation You will be as...

Page 14: ...me Conf P oFF x 12 Read ID number x 223 Set power off time Conf P oFF x 22 Read min alarm rail AL AL Lo x x 224 Logger Read data of CYCL Logger x 23 Read max alarm rail AL AL Hi x 225 Logger Read cycl...

Page 15: ...e measuring range by re plugging the tube to pressure port A Please note that all values are displayed as positive values No minus sign will be shown Example for HND PS02 For tube connection B the mea...

Page 16: ...Sensor defective Return to manufacturer for repair Err 2 Measured value below allowable range Check pressure not within sensor range measuring value to low Sensor defective Return to manufacturer for...

Page 17: ...ensor Accuracy typ 0 1 F S at nominal temperature FASt and P dET 0 5 FS Measuring rate slow 4 meas sec ConF Rate slow fast 1000 meas sec ConF Rate FASt and P dEt Nominal temperature 25 C Sensor All se...

Page 18: ...consumption Slow measuring rate 1 6mA Fast measuring rate 7 0mA Low Power Logger 0 3mA for cycle time 30s without interface communication active and no alarm horn sounding up to 0 8mA at cycle time 1s...

Page 19: ...HND P210 HND P210 K08 0422 page 19 11 Order Codes Order no Housing design HND P210 1 x pressure sensor input with additional functions...

Page 20: ...S06 0 2000 mbar abs HND PS07 0 7 00 bar abs 10 mbar max 10 bar abs HND PS08 0 400 0 mbar rel 0 1 mbar max 2 bar rel 0 70 C External threads G Edelstahl HND PS19 0 1000 mbar rel 1 mbar max 5 bar rel HN...

Page 21: ...olated HND Z033 Adapter RS232 converter on USB interface HND Z034 Windows software for setting and data read and print out of instruments of the HND series with logger function BUS S20M Software for r...

Page 22: ...ds noted below EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements Part 1 General requirements EN IEC 63000 2018 Technical documentation for the assessment...

Reviews: