8.9  Error and System Messages 

Display

 

Meaning

 

What to do?

 

 

Low battery power, device will only continue 
operation for a short period of time 

Replace battery 

 

Battery empty 

Replace battery 

Mains operation without battery: wrong 
voltage 

Check power supply, replace it when 
necessary 

 

or 

Err.9

 

No sensor connected 

Switch off device and connect sensor

Connected sensor or device defective 

If second sensor available, check if 
device is ok. Return defective 
device/sensor to manufacturer for repair

Value extremely out of measuring range 

Check: pressure not within sensor 
range?

No display or 

confused 

characters, 

device does 

not react on 

keypress 

Battery empty 

 

Mains operation: wrong  

 

 

 

 

 

Err.1

 

Measured value above allowable range 

Check: pressure not within sensor 
range? 
-> measuring value to high 

Sensor defective 

Return to manufacturer for repair 

Err.2

 

Measured value below allowable range 

Check: pressure not within sensor 
range? 
-> measuring value to low 

Sensor defective 

Return to manufacturer for repair 

Err.3 

Display range overflow 

Check: value above 19999->to high to 
be displayed

Err.4

 

Display range underflow 

Check: value below -19999 (Tara?) ->to 
low!

Err.11 

Value could not be calculated

Choose different unit 

Calculation overflow happened

Choose different unit 

Err.7

 

System error 

Return to manufacturer for repair 

---- 

Sensor not present/recognised 

Reconnect sensor, during logging: stop 
the logger and restart it 

Could not calculate value 

Suitable sensor/unit combination 
necessary

 
 

8.10 Calibration Services 

Calibration certificates – DKD-certificates – other certificates: 
If device should be certificated for its accuracy, it is the best solution to return it 
with the referring sensors to the manufacturer. 
Only the manufacturer is capable to do efficient recalibration if necessary, to get 
results of highest accuracy! 

 

Summary of Contents for HND-P210

Page 1: ...Operating Instructions for Hand held Pressure Measuring Devices for External Pressure Sensors Model HND P210...

Page 2: ...r Level Display Unit m 11 8 6 Operation Of Logger 11 8 7 The Serial Interface 14 8 8 Pressure Connection to The Sensors 15 8 9 Error and System Messages 16 8 10 Calibration Services 16 9 Maintenance 1...

Page 3: ...iced by persons familiar with these operating instructions and in accordance with local regulations applying to Health Safety and prevention of accidents When used in machines the measuring unit shoul...

Page 4: ...eration With Power Supply Warning When using a power supply please note that operating voltage has to be 10 5 to 12 VDC Do not apply overvoltage Cheap 12 V power supplies often have excessive no load...

Page 5: ...e the device temperature has adjusted to the ambient temperature before trying a new start up 4 If device is to be connected to other devices e g via serial interface the circuitry has to be designed...

Page 6: ...Furthermore the kind of measuring is displayed Slo Fast P det p r t 2 1 and Corr if a Sensor with activated offset or scale correction is connected After changing the battery the clock setting menu is...

Page 7: ...tment press button 3 for approx 5 seconds Auto Null will be displayed shortly The adjustment is done via the OFFSET value of the sensor referring configuration menu Note A zero point adjustment can on...

Page 8: ...mbar Enter desired value at which the analogue output potential should be 1 V if Out dAC Set CAL Set Calibration Adjustment of Sensor OFFS Sensordep e g 5 00 5 00 mbar The offset of sensor will be dis...

Page 9: ...Application with logger function Measuring of short pressure peaks or fast changing pressures with a resolution of 1ms The cyclic logger function records the arithmetic mean value the highest and the...

Page 10: ...rface converter To get access to each device the base addresses of the devices have to be different For example choose 01 for the first 11 for the second device and so on See also chapter 8 7 The Seri...

Page 11: ...tore 2 seconds The logger records 3 measurement results each time current or mean value depending on logger setting see below min peak and max peak of sensor Min and max peak are the minimum resp the...

Page 12: ...etween the different values referring to the measurement also is done by pressing Changing the measurement is done by pressing the keys or 8 6 2 Func CYCL Automatic Recording With Selectable Logger Cy...

Page 13: ...ording Do not stop the recording The selection can be made by key 2 and key 5 Quit key 6 enters the choice Note If you try to switch off the instrument in the cyclic recording operation You will be as...

Page 14: ...me Conf P oFF x 12 Read ID number x 223 Set power off time Conf P oFF x 22 Read min alarm rail AL AL Lo x x 224 Logger Read data of CYCL Logger x 23 Read max alarm rail AL AL Hi x 225 Logger Read cycl...

Page 15: ...e measuring range by re plugging the tube to pressure port A Please note that all values are displayed as positive values No minus sign will be shown Example for HND PS02 For tube connection B the mea...

Page 16: ...Sensor defective Return to manufacturer for repair Err 2 Measured value below allowable range Check pressure not within sensor range measuring value to low Sensor defective Return to manufacturer for...

Page 17: ...ensor Accuracy typ 0 1 F S at nominal temperature FASt and P dET 0 5 FS Measuring rate slow 4 meas sec ConF Rate slow fast 1000 meas sec ConF Rate FASt and P dEt Nominal temperature 25 C Sensor All se...

Page 18: ...consumption Slow measuring rate 1 6mA Fast measuring rate 7 0mA Low Power Logger 0 3mA for cycle time 30s without interface communication active and no alarm horn sounding up to 0 8mA at cycle time 1s...

Page 19: ...HND P210 HND P210 K08 0422 page 19 11 Order Codes Order no Housing design HND P210 1 x pressure sensor input with additional functions...

Page 20: ...S06 0 2000 mbar abs HND PS07 0 7 00 bar abs 10 mbar max 10 bar abs HND PS08 0 400 0 mbar rel 0 1 mbar max 2 bar rel 0 70 C External threads G Edelstahl HND PS19 0 1000 mbar rel 1 mbar max 5 bar rel HN...

Page 21: ...olated HND Z033 Adapter RS232 converter on USB interface HND Z034 Windows software for setting and data read and print out of instruments of the HND series with logger function BUS S20M Software for r...

Page 22: ...ds noted below EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements Part 1 General requirements EN IEC 63000 2018 Technical documentation for the assessment...

Reviews: