Kobold HND-P210 Operating Instructions Manual Download Page 5

 

 

HND-P210 

HND-P210 K08/0422 

 

page 5 

7.  Operation 

7.1  Safety Requirements 

This  device  has  been  designed  and  tested  in  accordance  with  the  safety 
regulations  for  electronic  devices.  However,  its  trouble-free  operation  and 
reliability cannot be guaranteed unless the standard safety measures and special 
safety advises given in this manual will be adhered to when using the device.

 

1.  Trouble-free operation and reliability of the device can only be  guaranteed if 

the device is not subjected to any other climatic conditions than those stated 
see chapter 

10 Technical Information

.

 

2.  Device  and  sensors  have  to  be  handled  with  care  (don’t  throw,  hit,  etc.). 

Protect plugs and sockets from soiling.

 

3.  If the device is transported from a cold to a warm environment condensation 

may cause in a failure of the function. In such a case make sure the device 
temperature  has  adjusted  to  the  ambient  temperature  before  trying  a  new 
start-up.

 

4.  If  device  is  to  be  connected  to  other  devices  (e.g.  via  serial  interface)  the 

circuitry  has  to  be  designed  most  carefully.  Internal  connection  in  third  party 
devices  (e.g.  connection  GND  and  earth)  may  result  in  not-permissible 
voltages impairing or destroying the device or another device connected.

 

Warning: If device is operated with a defective mains power 
supply (e.g. short circuit from mains voltage to output voltage) 
this may result in hazardous voltages at the device (e.g. at sensor 
socket or interface).

 

5.  If  there  is  a  risk  whatsoever  involved  in  running  it,  the  device  has  to  be 

switched off immediately and to be marked accordingly to avoid re-starting.

 

 Operator safety may be a risk if:

 

  there is visible damage to the device 

  the device is not working as specified 

  the device has been stored under unsuitable conditions for a longer period 

of time. 

In case of doubt, please return device to manufacturer for repair or maintenance.

 

 
 

Summary of Contents for HND-P210

Page 1: ...Operating Instructions for Hand held Pressure Measuring Devices for External Pressure Sensors Model HND P210...

Page 2: ...r Level Display Unit m 11 8 6 Operation Of Logger 11 8 7 The Serial Interface 14 8 8 Pressure Connection to The Sensors 15 8 9 Error and System Messages 16 8 10 Calibration Services 16 9 Maintenance 1...

Page 3: ...iced by persons familiar with these operating instructions and in accordance with local regulations applying to Health Safety and prevention of accidents When used in machines the measuring unit shoul...

Page 4: ...eration With Power Supply Warning When using a power supply please note that operating voltage has to be 10 5 to 12 VDC Do not apply overvoltage Cheap 12 V power supplies often have excessive no load...

Page 5: ...e the device temperature has adjusted to the ambient temperature before trying a new start up 4 If device is to be connected to other devices e g via serial interface the circuitry has to be designed...

Page 6: ...Furthermore the kind of measuring is displayed Slo Fast P det p r t 2 1 and Corr if a Sensor with activated offset or scale correction is connected After changing the battery the clock setting menu is...

Page 7: ...tment press button 3 for approx 5 seconds Auto Null will be displayed shortly The adjustment is done via the OFFSET value of the sensor referring configuration menu Note A zero point adjustment can on...

Page 8: ...mbar Enter desired value at which the analogue output potential should be 1 V if Out dAC Set CAL Set Calibration Adjustment of Sensor OFFS Sensordep e g 5 00 5 00 mbar The offset of sensor will be dis...

Page 9: ...Application with logger function Measuring of short pressure peaks or fast changing pressures with a resolution of 1ms The cyclic logger function records the arithmetic mean value the highest and the...

Page 10: ...rface converter To get access to each device the base addresses of the devices have to be different For example choose 01 for the first 11 for the second device and so on See also chapter 8 7 The Seri...

Page 11: ...tore 2 seconds The logger records 3 measurement results each time current or mean value depending on logger setting see below min peak and max peak of sensor Min and max peak are the minimum resp the...

Page 12: ...etween the different values referring to the measurement also is done by pressing Changing the measurement is done by pressing the keys or 8 6 2 Func CYCL Automatic Recording With Selectable Logger Cy...

Page 13: ...ording Do not stop the recording The selection can be made by key 2 and key 5 Quit key 6 enters the choice Note If you try to switch off the instrument in the cyclic recording operation You will be as...

Page 14: ...me Conf P oFF x 12 Read ID number x 223 Set power off time Conf P oFF x 22 Read min alarm rail AL AL Lo x x 224 Logger Read data of CYCL Logger x 23 Read max alarm rail AL AL Hi x 225 Logger Read cycl...

Page 15: ...e measuring range by re plugging the tube to pressure port A Please note that all values are displayed as positive values No minus sign will be shown Example for HND PS02 For tube connection B the mea...

Page 16: ...Sensor defective Return to manufacturer for repair Err 2 Measured value below allowable range Check pressure not within sensor range measuring value to low Sensor defective Return to manufacturer for...

Page 17: ...ensor Accuracy typ 0 1 F S at nominal temperature FASt and P dET 0 5 FS Measuring rate slow 4 meas sec ConF Rate slow fast 1000 meas sec ConF Rate FASt and P dEt Nominal temperature 25 C Sensor All se...

Page 18: ...consumption Slow measuring rate 1 6mA Fast measuring rate 7 0mA Low Power Logger 0 3mA for cycle time 30s without interface communication active and no alarm horn sounding up to 0 8mA at cycle time 1s...

Page 19: ...HND P210 HND P210 K08 0422 page 19 11 Order Codes Order no Housing design HND P210 1 x pressure sensor input with additional functions...

Page 20: ...S06 0 2000 mbar abs HND PS07 0 7 00 bar abs 10 mbar max 10 bar abs HND PS08 0 400 0 mbar rel 0 1 mbar max 2 bar rel 0 70 C External threads G Edelstahl HND PS19 0 1000 mbar rel 1 mbar max 5 bar rel HN...

Page 21: ...olated HND Z033 Adapter RS232 converter on USB interface HND Z034 Windows software for setting and data read and print out of instruments of the HND series with logger function BUS S20M Software for r...

Page 22: ...ds noted below EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements Part 1 General requirements EN IEC 63000 2018 Technical documentation for the assessment...

Reviews: