Density calibration
Enter customer's
password
Measurement
hot Medium
Go to subitem measuremnt
hot medium
Select "yes" to measure
actual frequency and
temperature
Finish density
calibration
End
Measured values
hot medium
confirm the measured
temperature and frequency
Enter te density of your
medium
Single point density
calibration without
manufacturer calibration
Warm up of sensor and
medium to operation
temperature
15 minutes minimum waiting
and operationg time
Determination of
medium's density
Determination of medium
density by external
measurement
(Quality of medium's density
determination should be 1g/l)
DENSITY
Density on
Switch on density
measurment
(herewith volume flow
meeasuremnt is also
possible)
Switch over to functional
class density
Single Point Density Calibration