Page 76 of 105
OPERATING MANUAL HPC & UMC4
14.9.2 Status output 1 assignment
This function allows the operator to define to which event the status output is to be assigned. The most
general assignment is the alarm assignment because all set limiting values and the self-test function are
then monitored via the status output.
After selecting the
Status output assignment
function, press
„Enter“ to display the current assignment.
Output 1 assigned to
[Alarm]
As mentioned in Section 13.4.3.1 Selection window/make a selection, the operator can choose between
the following settings:
Flow direction recognition
o
Forward flow
o
Reverse flow
Limiting values:
o
MIN QM
o
MAX QM
o
MIN density
o
MAX density
All limiting values and error detection
o
Alarm
Pulse output 2 for custody transfer operations
o
IMP2 90°,
When selecting the IMP2 90° setting, a second pulse output will be realized via the status output that can
be used for custody transfer operations.
o