17.3.2 Display of system error
System errors consist of the message text “system error” and a 5-digit number in hexadecimal code. The
meaning of the individual error codes is described in the following table. If several errors occur at the
same time, the hexadecimal sum of the individual errors will be displayed. The errors are coded in such a
way that the individual errors can be easily identified. The sums are unique.
Descriptor label
(never displayed)
Constant/
display
Description
SystemfehlerDiv0
0x00001
Arithmetical error/division by zero
SystemfehlerIntEEProm
0x00002
Transmitter data checksum is faulty; reinitialization is
necessary.
SystemfehlerPruefsumme
0x00004
Sensor data checksum is faulty.
SystemfehlerLeeresEEPROM
0x00008
Ext. EEPROM is present but empty (no content).
SystemfehlerEEPROM
0x00010
Value could not be stored/read out.
SystemfehlerPhase
0x00020
Phase measurement/mass flow is faulty.
SystemfehlerFrequenz
0x00040
Frequency measurement/density measurement is
faulty.
Systemfehler DSP Version
0x00080
DSP firmware is outdated (not adjusted to the transmit-
ter operating system)
SystemfehlerZeitkonstante
0x00100
Initialization of time constants failed.
SystemfehlerMesswert
0x00200
Faulty calculation of measured value
SystemfehlerParameter
0x00400
Settings are inconsistent.
SystemfehlerRAMPrüfsumme
0x00800
Defective main memory, inconsistent checksum (custo-
dy transfer operation)
SystemfehlerFlashPrüfsumme
0x01000
Defective program memory, inconsistent checksum
SystemfehlerDSPPrüfsumme
0x02000
Defective program memory, inconsistent checksum
SystemfehlerZähler
0x04000
Custody transfer operation: count differs from corre-
sponding back-up copy