OPERATING MANUAL HPC & UMC4
Page 93 of 105
14.14.3 Flow direction
This function allows the operator to define the flow direct
ion that the transmitter will evaluate. Only “for-
ward” should be selected so as to prevent reverse flow from being measured. The standard factory set-
ting is “forward & reverse.” After selecting the
Flow direction
function, press
„Enter“ to display the current
setting.
Flow direction
[forward]
As mentioned in Section 13.4.3.1 Selection window/make a selection the operator can choose between:
forward
reverse
forward & reverse
0%
20%
40%
60%
80%
100%
120%
Ou
tpu
t
(cu
rr
en
t,
p
u
lse,
fr
eq
u
en
cy
)
measured flow rate
Flow direction
forward flow
reverse flow
forward & reverse