OPERATING MANUAL TMU-W / UMC4
7.2.3.1
Unit of counters
After choosing the Unit of counters function and pressing “Enter”, the current forward and reverse
counter unit will be displayed:
Accumulation of:
[kg]
One of the following units can be selected.
Mass units:
g, kg, t, lbs, shton, lton and oz
Volume units:
m³, cm³, l, USG, UKG, USB, ft³, acf and floz
Programmable mass unit:
xxQM,
Programmable volume unit:
xxQV.
When the unit is changed, the counters will be reset to 0.00 automatically.
The volume unit only makes sense if the sensor has been calibrated for density measurement. Press “Enter” to
confirm and save the selection. Forward and reverse counters will now show the selected unit.
The valence of the programmable units are defined by the settings of the flow units described in sections 7.2.5.2,
“Factor mass flow QM programmable unit” on page 62 and 7.2.5.8, “Factor volume flow QV programmable unit”
on page 65.
7.2.3.2
Reset counters
To reset the totalizing counters, you need to toggle to [yes]. Forward and reverse counters will be reset at the
same time (0.00).
Reset counters
[no]
By pressing “Esc” or toggling to [no] the operator can cancel the current action without changing the counter
readings.