DON-H
page 34
DON-H
01/0721
15.1.5
Troubleshooting
No changes may be made to products that are operated in connection with
potentially explosive areas. Repairs to the product may only be carried out by
specially trained and authorized specialists.
15.1.6
Disposal
The packaging and used parts must be disposed of in accordance with the
regulations of the country in which the product is installed.
15.2 ATEX version Exi (electronics options xA)
While evaluating the product, following Norms were considered:
a) EN IEC 60079-0: 2018 Ed. 7 Hazardous areas - Part 0: Equipment -
General requirements
b) IEC 60079-11: 2011 Ed. 6 + Corr. 2012 / EN 60079-11: 2012 Hazardous
areas - Part 11: Equipment protection by intrinsic safety "i"
c) EN ISO 80079-36 2016 Explosive atmospheres - Part 36: Non-electrical
devices for use in explosive atmospheres - Basics and requirements.
d) EN ISO 80079-37 2016 Explosive atmospheres - Part 37: Non-electrical
equipment for explosive atmospheres – Non electrical type of protection
constructional safety "c", control of ignition source "b", liquid immersion "k"
15.2.1
General information on explosion protection
The electrical connection is made via a plug in a separate inductively coupled
electronics compartment - exclusively in intrinsically safe systems. The
electronics option HA is combined with the electronics models ZOK-E * M
(compact version) or ZOK-E * K (remote electronics) in accordance with LOM 16
ATEX 2062 to form the electronics options 1A, 2A, 3A and 5A.
Permitted use
The intrinsically safe flow meter can be used as follows:
In Zone 2 (Gas-Ex, Category 3G, EPL Gc) in explosion groups IIA, IIB and
IIC.
The requirements for simple electrical equipment in intrinsically safe circuits
in Zone 1 according to EN 60079-11 are met for various versions.
The qualification with regard to the surface temperature is T4 / T3; for all
gases, vapours and fog with an ignition temperature ≥ 135 ° C / ≥ 200 ° C, the
products are not an ignition source.
The permissible ambient temperature range is from -20 ° C ≤ Ta ≤ 80 ° C.
In accordance with the installation regulations, housing must be grounded using
the earthing clamps. The maximum connection cross-section is 4 mm
2
.
Summary of Contents for DON-05
Page 1: ...Operating Instructions for Oval Wheel Flowmeter High pressure Model DON H...
Page 9: ...DON H DON H 01 0721 page 9 Note The electronic option H0 is available with reed switch output...
Page 43: ...DON H DON H 01 0721 page 43 17 Statement of Conformity...
Page 44: ...DON H page 44 DON H 01 0721...
Page 45: ...DON H DON H 01 0721 page 45...
Page 46: ...DON H page 46 DON H 01 0721...
Page 47: ...DON H DON H 01 0721 page 47...
Page 48: ...DON H page 48 DON H 01 0721 18 State of safeness...