62
14 Specifications
Compressed air
Compressed air quality to ISO 8573-1:2001
Quality class 5.3.3
Solid contaminants
Class 5 (max. 40 µm, max. 10 mg/m
3
)
Water content
... for temperatures ≥ 15 °C:
... for temperatures 5 ... 15 °C:
Oil content
Class 4
With operating temperatures > 15 °C
a pressure dew point of max 3 °C is permitted
Class 3
Pressure dew point -20 °C (or below)
Class 3 (max. 1 mg/m
3
)
Permitted pressure range
4
*
… 10 bar
Operating pressure
for sensor lock gate
4
*
… 7 bar
Pressure monitoring
Automatic monitoring, message
Connection
G 1/4" female thread
Air consumption
Max. 300 L/min during probe movement
Min. air temperature
5 °C
Rinse water
Filtered 100 µm
Permitted pressure range
2 … 6 bar
Temperature range
5 … 65 °C
Pressure monitoring
Automatic monitoring, message
Connection
G 1/4" female thread /
G 3/4" male thread
Media adapter
Three ports for metering pump
• Port I and II:
Calibration buffer
• Port III:
Cleaning agent
Material
See bill of material
Ingress protection
IP 65
Mounting
Wall or pipe mounting (Option)
* Increased minimum pressure of 5 bar required for retractable fitting
in the case of high process pressure or difficult process media
Summary of Contents for Unical 9000
Page 16: ...16 ...
Page 68: ...68 Pneumatic Diagram of Unical 9000 ...
Page 69: ...69 Pneumatic Diagram of Media Connection ...
Page 79: ......