Stratos Multi E401N
18
2.4.1 Nameplate
1
3
7
6
8
2
4
9
10
11
5
1 Name of manufacturer
6 Protection class II
2 Product name
7 Power supply
3 CE marking
8 Manufacturer’s address with barcode
4 Special conditions: Read the user man-
ual, observe the specifications, and fol-
low the instructions in the safety guide.
9 Permissible ambient temperature, protection class
10 Product number/Serial number/Production year and week
5 Reminder to read the documentation
11 Model designation
You can view the device type, serial number, and firmware, hardware, and bootloader versions of your
device in the Diagnostics menu:
Menu
Diagnostics
Device Information
2.5 Symbols and Markings on the Display
Function check in accordance with NAMUR NE 107
Wrench symbol on orange background
The NAMUR “HOLD” contact is active. Current outputs as configured:
Currently measured value: The currently measured value appears at the current output.
Last measured value: The last measured value is held at the current output.
Fixed value: The current output supplies a fixed value.
Out of specification in accordance with NAMUR NE 107
Black question mark on yellow background
The NAMUR “Out of Specification” contact is active.
Error message:
Diagnostics
Message List
Failure in accordance with NAMUR NE 107
Flashing black cross symbol on red background
The NAMUR “Failure” contact is active.
Error message:
Diagnostics
Message List
Maintenance required in accordance with NAMUR NE 107
Oil can symbol on blue background
The NAMUR “Maintenance Required” contact is active.
Error message:
Diagnostics
Message List
The device is in calibration mode. Function check (HOLD) is active.
The device is in maintenance mode. Function check (HOLD) is active.
The device is in parameter setting mode. Function check (HOLD) is active.
The device is in diagnostics mode.