Stratos Multi E401N
114
6.4.6 Calibration Mode: Data Entry
Entry of values for the cell constant of a sensor, related to 25 °C/77 °F.
Calibration Procedure
Calibration
[I] [II] … Cond
When you access calibration, the calibration values selected in the calibration presettings are used.
These settings can still be changed in the Calibration menu. If you do not want to perform calibration,
use the left
softkey
to go back one level and abort the calibration.
01.
Select “Data Entry” calibration mode and press
enter
to confirm.
02.
Remove the sensor and install the premeasured sensor.
03.
Continue with the
right softkey: Next
.
04.
Enter the cell constant of the premeasured sensor.
Result
✔
By pressing the
right softkey: Adjust
, the calibration values obtained during calibration for
calculation of the process variables are applied to the device. When using a Memosens sensors,
the calibration values are saved in the sensor.
6.4.7 Calibration Mode: Temperature
Adjustment of Temperature Probe
This function is used to adjust the individual temperature probe tolerances or cable lengths for the
purpose of increasing the accuracy of the temperature measurement.
The adjustment requires an accurate measurement of the process temperature using a calibrated
reference thermometer. The measurement error of the reference thermometer should be less than 0.1
K. Adjustment without an accurate measurement may result in falsification of the displayed measured
value.
When using Memosens sensors, the adjustment value is saved in the sensor.
Calibration Procedure
Calibration
[I] [II] [Sensor]
Adjust Temperature Probe
01.
Select “Temperature” calibration mode and press
enter
to confirm.
02.
Enter the measured process temperature and press
enter
to confirm.
✓
The temperature offset is displayed.
03.
Adjust the temperature probe with the
right softkey: Save
.