Stratos Multi E401N
164
Saturation
1)
0.000 ... 150.0 %
Concentration
1)
(dissolved oxygen)
0000 ... 9999 µg/l /10.00 ... 20.00 mg/l
0000 ... 9999 ppb/10.00 ... 20.00 ppm
Volume concentration in gas
000.0 ... 9999 ppm/1.000 ... 50.00 vol%
Trace sensor “001” (analog)
Saturation
1)
0.000 ... 150.0 %
Concentration
1)
(dissolved oxygen)
000.0 ... 9999 µg/l /10.00 ... 20.00 mg/l
000.0 ... 9999 ppb/10.00 ... 20.00 ppm
Volume concentration in gas
000.0 ... 9999 ppm/1.000 ... 50.00 vol%
Input correction
Pressure correction
0.000 ... 9999 bar/999.9 kPa/145.0 psi (adjustable)
Manual or external (via current input 0(4) ... 20 mA)
Salinity correction
0.0 ... 45.0 g/kg
Calibration and adjustment
Automatic calibration in air-saturated water
Automatic calibration in air
Product calibration, saturation (with offset on SE 740)
Zero correction
Temperature probe adjustment
Calibration ranges
Standard sensor
Zero point
± 2 nA
Slope
25 ... 130 nA (at 25 °C/77 °F, 1013 mbar)
Trace sensor “01”
Zero point
± 2 nA
Slope
200 ... 550 nA (at 25 °C/77 °F, 1013 mbar)
Trace sensor “001”
Zero point
± 3 nA
Slope
2000 ... 9000 nA (at 25 °C/77 °F, 1013 mbar)
Calibration timer
0000 ... 9999 h
1)
For temperature range -10 ... 80 °C/14 ... 176 °F