108
Parameter Setting: Installation
Configuration of Uniclean 900 Functions
Installation
Default setting
Adjustable parameters
•
Measurement procedures Continuous
(Continuous / Short-time)
•
Ext. control (DCS)
(Polarity / Output settings)
- Signal level of inputs
DCS (36 ... 39)
M/S (42, 43)
A/M (40, 41)
Active: 10 ... 30 V
(Active: 10 ... 30 V / active < 2V)
- Signal level of outputs
DCS (31 ... 34)
N/O
(N/O / N/C)
•
Sensor detection
Off
On
•
Immersion lock
Off
Sensocheck Glass el
•
Access manual control
Access code for manual
control (Maintenance menu)
Default: 2958
•
Probe
- Probe type
Ceramat
(SensoGate, Other)
- Max. move time
0015 sec
- Sealing water
Off
(On)
- Cavity rinsing
Off
(Off, Interval, Continuous)
- Check interval
Off
On (Check after travels)
- Maintenance interval
Off
On (Maintenance after travels)
•
Media adapter
- Metering pump
Off*
“On” or “Off”
- Medium:
---
(e.g. “Cleaning agent A”)
- Displaced volume
50 ml
(25 / 50 / 75 / 100 ml)
- Residual volume
250 ml
(0 / 250 / 500 ml)
•
Additional media
- Additional medium 1
Off
(On - then enter designation)
- Additional medium 2
Off
(On - then enter designation)
•
Start-up
No
Yes/No
•
System forecast
Off
Off / On: Monitors the probe travels for
predictive maintenance of Ceramat und
SensoGate
See "System Forecast (Ceramat,
SensoGate)" on page 88 for description.
*Automatic adjustment by “Plug & Play” in: System control / Factory setting Uniclean
Parameter setting:
Installation
Summary of Contents for Protos 3400
Page 12: ...12 ...
Page 160: ...160 ...
Page 177: ...177 Appendix Buffer Table Mettler Toledo ...
Page 179: ...179 Appendix Buffer Table DIN 19267 ...
Page 181: ...181 Appendix Buffer Table Techn Buffers to NIST ...
Page 182: ...182 Appendix Buffer Table Hamilton ...
Page 183: ...183 Appendix Buffer Table Kraft ...
Page 184: ...184 Appendix Buffer Table Hamilton A ...
Page 185: ...185 Appendix Buffer Table Hamilton B ...
Page 186: ...186 Appendix Buffer Table HACH ...
Page 206: ...206 Index ...
Page 207: ...207 Index ...