190
Parameter setting
Activated from measuring mode: Press menu key to select menu.
Select parameter setting using arrow keys, confirm with enter.
Administrator level
Access to all functions, also passcode setting. Releasing or
blocking functions for access from the Operator level.
Operator level
Access to all functions which have been released at the
Administrator level. Blocked functions are displayed in gray and
cannot be edited.
Viewing level
Only display, no editing possible!
Menu selection
Return to meas
Select:
[enter]
Lingua
Parameter setting
Return
Viewing level
Administrator level (All Data) adm
(Operation Data) opl
(All Data) view
Operator level
System control
Parameter sets
• Load
• Save
Overview of Parameter Setting
2 parameter sets (A,B) are available in the analyzer.
The currently active parameter set is read on the display.
Parameter sets contain all settings except:
Sensor type, Options, System control settings
Up to 5 parameter sets (1, 2, 3, 4, 5) are available when a
SmartMedia card (Option) is used.
Memory card (Option)
• Record logbook
• Register recorder
• Decimal separator
• Card full
• Format
Menu only appears with SmartMedia Card inserted.
Make sure that it is a memory card,
not an update card.
Commercially available SmartMedia cards must be formatted in
the analyzer before they can be used as memory card.
Copy configuration
The complete configuration of an analyzer can be written on a
SmartMedia card. This allows transferring all device settings to
other devices with identical equipment (exception: options and
passcodes).
Function control matrix
• Input OK2
• Left softkey
• Right softkey
Selecting the control element for the following functions:
- Parameter set selection
- KI recorder (start/stop)
- Favorites menu (selected diagnostics functions)
- Unical (fully automated probe controller)
Time/date
Point of meas description
Release of options
Software update
Logbook
Buffer table
Factory setting
Passcode entry
Selecting the display format, entry
Can be called from the diagnostics menu.
A TAN is required to release an Option.
Software update from SmartMedia card (update card)
Selecting events to be recorded
Entering own buffer set for automatic calibration
Resetting all parameters to factory setting
Editing the passcodes
25.6 °C
7.00 pH
25.6 °C
7.00 pH
Overview of Parameter Setting
Summary of Contents for Protos 3400
Page 12: ...12 ...
Page 160: ...160 ...
Page 177: ...177 Appendix Buffer Table Mettler Toledo ...
Page 179: ...179 Appendix Buffer Table DIN 19267 ...
Page 181: ...181 Appendix Buffer Table Techn Buffers to NIST ...
Page 182: ...182 Appendix Buffer Table Hamilton ...
Page 183: ...183 Appendix Buffer Table Kraft ...
Page 184: ...184 Appendix Buffer Table Hamilton A ...
Page 185: ...185 Appendix Buffer Table Hamilton B ...
Page 186: ...186 Appendix Buffer Table HACH ...
Page 206: ...206 Index ...
Page 207: ...207 Index ...