35
Calcheck (Administrator)
20.1 °C
Abort
7.00 pH
Meas distance max +02.00 pH
OK
Off
Failure
Maint. request
Message
Parameter Setting of PHU 3400(X)-110
Note: HOLD mode active
Menu
Module PHU 3400-110 (Administrator)
20.1 °C
Return
TC process medium
Sensor data
Cal preset values
7.00 pH
ORP/rH value
Delta function
Input filter
Sensor data (Administrator)
20.1 °C
Abort
Sensoface
7.00 pH
Temperature detection
Display
Display
Parameter selection
Sensor type
Slope
Zero
Block
Block
OK
Standard
Standard
(Pt100)
Off
On
Sensocheck Ref el
Sensor data
Sensor data are preset depending on
the sensor type.
Gray display lines cannot be edited.
Sensoface provides information on
the sensor condition (evaluating the
sensor data).
Great deviations are signaled.
Sensocheck can be switched off.
The following parameters are
monitored: Slope, zero, reference
impedance (for ISFET sensors),
glass impedance (pH electrodes),
response time, CalCheck.
The tolerance limits are displayed in
gray.
Limit violations can be signaled by a
message (either "failure" or
"maintenance request").
Sensor monitoring details
20.1 °C
Abort
Sensocheck Glass el
7.00 pH
Zero
Sensocheck Ref el
(Auto)
Slope
Response time
Calcheck
OK
(Auto)
Sensor monitoring details
With “Auto” selected, the limits
calculated by the analyzer are used
(displayed in gray).
With “Sensor monitoring Individual”,
the tolerance limits for the parameters
are displayed in black and can be set.
(Auto)
(Auto)
(Auto)
(Auto)
Summary of Contents for Protos 3400
Page 12: ...12 ...
Page 160: ...160 ...
Page 177: ...177 Appendix Buffer Table Mettler Toledo ...
Page 179: ...179 Appendix Buffer Table DIN 19267 ...
Page 181: ...181 Appendix Buffer Table Techn Buffers to NIST ...
Page 182: ...182 Appendix Buffer Table Hamilton ...
Page 183: ...183 Appendix Buffer Table Kraft ...
Page 184: ...184 Appendix Buffer Table Hamilton A ...
Page 185: ...185 Appendix Buffer Table Hamilton B ...
Page 186: ...186 Appendix Buffer Table HACH ...
Page 206: ...206 Index ...
Page 207: ...207 Index ...