32
pH
Oxy Cond
ORP
Data logger full
The Data Logger
The meter provides a data logger. Prior to use, it must be configured and then
activated. You can choose from the following logger types:
• Shot (manual logging by pressing the Save value softkey)
• Interval (time-controlled logging at a fixed interval)
• Difference (signal-controlled logging of measured variable and temperature)
• Intv+Diff (combined time- and signal-controlled logging)
• Limit value (combined time- and threshold-controlled logging)
The data logger records up to 10 000 entries, which can be assigned to different
points of measurement (TAGs) and notes. The following data will be recorded:
meas. point, note, sensor ID, serial number of sensor (Memosens), primary value,
temperature, time stamp, device status.
The Paraly SW 112 software allows convenient management of the data logger.
It is always the currently selected process variable which is recorded.
Data logger activated
Number of measured values stored
Circular
recording
Start/Stop logger
(for softkey configuration see page 14)
If logger
full
Menu
Stop logger
Display: Icons related to the data logger
Temperature
Process
variable
Data Logger