Sensor Network Diagram (pH and Oxy only)
Provide single-glance information on the following parameters of the connected
sensor:
• Slope
• Zero point (operating point for Memosens ISFET)
• Sensocheck (pH) or leakage current (ISFET and Oxy)
• Response time
• Calibration timer
• Wear (Memosens)
Parameters which cannot be checked are shown as inactive (gray) and are set to
100% (e.g. Sensocheck for analog sensors).
The parameter values should lie between the outer (100%) and inner (50%) hexa-
gon. When a value enters the inner hexagon (<50%), the corresponding caption text
flashes red (see example).
Information
Help
Back
Network Diagram
pH
Response time
Sensocheck
Zero point
Calibration timer
Wear
Slope
Example: Network diagram of a digital pH sensor (Memosens)