![Knick PORTAVO 904X PH User Manual Download Page 26](http://html.mh-extra.com/html/knick/portavo-904x-ph/portavo-904x-ph_user-manual_1980164026.webp)
26
The Data Logger
The meter provides a data logger.
Prior to use
, it must be configured and then
activated. You can choose from the following logger types:
• DIFF (signal-controlled logging of measured variable and temperature)
• INT (time-controlled logging at a fixed interval)
• DIFF+INT (combined time- and signal-controlled logging)
• SHOT (manual logging by pressing the
STO
key)
The data logger records up to 5000 entries and saves them in a circular buffer.
Already existing entries will be overwritten.
The following data are recorded: primary value, temperature, time stamp and device
status.
Option 001 SOP can be used to set up an access lock for the data logger, which in the
absence of an access code allows only logger data to be displayed (see page 35).
The Paraly SW 112 PC software allows convenient management of the data logger.
It is always the currently selected process variable which is recorded. The “STO” icon
and the memory address is displayed briefly to indicate that an entry is being saved.
Data Logger
RCL
:
Saved mea-
sured values
are read.
Logger type
Date
Data logger activated
Primary
process
variable
Temperature
Display: Icons Related to the Data Logger
STO:
Measured
value is saved.
Memory address (0026)
Next value
Time
Summary of Contents for PORTAVO 904X PH
Page 2: ......
Page 50: ...50 pH Buffer Tables 01 Mettler Toledo Nominal values in bold C pH ...
Page 51: ...51 pH 02 Knick CaliMat Nominal values in bold C pH Buffer Tables ...
Page 52: ...52 pH 03 Ciba 94 Nominal values 2 06 4 00 7 00 10 00 C pH 1 extrapolated Buffer Tables ...
Page 53: ...53 pH 04 Technical Buffers to NIST Nominal values in bold C pH 1 values added Buffer Tables ...
Page 55: ...55 pH 06 HACH Nominal values 4 01 7 00 10 01 0 02 at 25 C C pH Buffer Tables ...
Page 56: ...56 pH 07 WTW Technical Buffers Nominal values in bold C pH Buffer Tables ...
Page 57: ...57 pH 08 Hamilton Nominal values in bold C pH Buffer Tables ...
Page 58: ...58 pH Buffer Tables 09 Reagecon Nominal values in bold C pH 1 values added ...
Page 59: ...59 pH Buffer Tables 10 DIN 19267 Nominal values in bold C pH 1 extrapolated ...
Page 66: ...66 V Value added features 8 Viewing recorded data 32 Viewing the logger data 32 Index ...
Page 67: ......