Measurement
cal
CAL
CAL SOP
Use
to set the required calibration method
(CAL SOP).
cal
Perform the selected calibration
(see Paraly SW 112 PC software for description).
The meter then automatically returns to measuring mode.
CAL SOP Calibration
(Option 001 SOP, must have been configured in Paraly SW 112
PC software)
In the Paraly SW 112 PC software, you specify which buffers are to be used in which
sequence. You can combine buffer solutions from different buffer sets. Please note
that the minimum distance allowed between two buffer solutions is Δ 2 pH.
SOP calibration allows you to:
• use up to 3 buffers for adjustment
• use a 4th point for verification (“verification buffer”)
specify a maximum deviation from the verification buffer
• use buffers from different buffer sets, including a “user buffer”.
Note:
To abort calibration, you can press
meas
at any time.
Calibration
Summary of Contents for PORTAVO 904X PH
Page 2: ......
Page 50: ...50 pH Buffer Tables 01 Mettler Toledo Nominal values in bold C pH ...
Page 51: ...51 pH 02 Knick CaliMat Nominal values in bold C pH Buffer Tables ...
Page 52: ...52 pH 03 Ciba 94 Nominal values 2 06 4 00 7 00 10 00 C pH 1 extrapolated Buffer Tables ...
Page 53: ...53 pH 04 Technical Buffers to NIST Nominal values in bold C pH 1 values added Buffer Tables ...
Page 55: ...55 pH 06 HACH Nominal values 4 01 7 00 10 01 0 02 at 25 C C pH Buffer Tables ...
Page 56: ...56 pH 07 WTW Technical Buffers Nominal values in bold C pH Buffer Tables ...
Page 57: ...57 pH 08 Hamilton Nominal values in bold C pH Buffer Tables ...
Page 58: ...58 pH Buffer Tables 09 Reagecon Nominal values in bold C pH 1 values added ...
Page 59: ...59 pH Buffer Tables 10 DIN 19267 Nominal values in bold C pH 1 extrapolated ...
Page 66: ...66 V Value added features 8 Viewing recorded data 32 Viewing the logger data 32 Index ...
Page 67: ......