![Knick A405B User Manual Download Page 90](http://html.mh-extra.com/html/knick/a405b/a405b_user-manual_1980154090.webp)
90
Specifications
Service functions
Current source
Current specifiable for output 1 and 2 (00.00 ... 22.00 mA)
Sensor monitor
Display of direct sensor signals (mV/temperature/operating time)
Relay test
Manual control of relay contacts
Device type
Selecting the measuring function
Data retention
Parameters, calibration data, logbook > 10 years (EEPROM)
Electrical safety
Protection against electric shock by protective separation of all extra-
low-voltage circuits against mains according to EN 61010-1
Explosion protection
(A405B)
see Control Drawing or www.knick.de
EMC
EN 61326
Emitted interference
Class B (residential environment)
Immunity to interference
Industrial environment
RoHS conformity
according to EC directive 2002/95/EC
Power supply
80 V (-15%) ... 230 (+10%) V AC ; approx. 15 VA ; 45 … 65 Hz
24 V (-15%) … 60 (+10%) V DC ; 10 W
Overvoltage category II, protection class II
Nominal operating conditions
Ambient temperature
-20 ... +55 °C /-4 ... +131 °F
Transport/Storage temperature
-30 ... +70 °C /-22 ... +158 °F
Relative humidity
10 ... 95% not condensing
Enclosure
Molded enclosure made of PBT/PC, glass fiber reinforced
Mounting
Wall, pipe/post or panel mounting
Color
Gray, RAL 7001
Ingress protection
IP 67 / NEMA 4X outdoor (with pressure compensation)
Flammability
UL 94 V-0
Dimensions
H 148 mm, W 148 mm, D 117 mm
Control panel cutout
138 mm x 138 mm to DIN 43 700
Weight
1.2 kg (1.6 kg incl. accessories and packaging)
Cable glands
3 knockouts for M20 x 1.5 cable glands
2 knockouts for NPT ½” or rigid metallic conduit
Connections
Terminals, conductor cross section max 2.5 mm
2
,