![Knick A405B User Manual Download Page 37](http://html.mh-extra.com/html/knick/a405b/a405b_user-manual_1980154037.webp)
37
Configuring the Current Output
Menu item
Action
Choices
Current range
Select 4-20 mA or 0-20 mA
range using
keys.
Press enter to confirm.
Process variable
Select using
keys.
Cond: Conductivity
TMP: Temperature
Press enter to confirm.
Then select characteristic (LIN/
biLIN/LOG).
Current start
Modify digit using
keys,
select next digit using
keys.
Press enter to confirm.
Entered value applies to
selected process variable/range.
If the adjusted range is
exceeded, the device automati-
cally switches to the next higher
range (Autorange)
Current end
Enter value using
keys.
Press enter to confirm.
Entered value applies to
selected process variable/range.
If the adjusted range is
exceeded, the device automati-
cally switches to the next higher
range (Autorange)
Assignment of measured values: Current start and current end
Output current
[mS/cm]
20
4
200
0
20
4
200
100
[mA]
Example 1: Measuring range 0 … 200 mS/cm
Example 2: Measuring range 100 … 200 mS/cm
Advantage: Higher resolution in range of interest
[mA]
Output current
100
[mS/cm]