65
Display
Action
Remark
Display of selected process
variable (here: mS/cm). Now
the device is in HOLD mode:
Reinstall the sensor and
check whether the measure-
ment is OK.
MEAS exits calibration,
REPEAT permits repetition.
With MEAS selected:
Press enter to exit
calibration.
Display of conductivity
and temperature, Senso-
face is active.
After end of calibration,
the outputs remain in
HOLD mode for a short
time.
After display of GOOD
BYE, the device automat-
ically returns to measur-
ing mode.
Note:
• Be sure to use known calibration solutions and the respective temperature-
corrected conductivity values (see “Calibration Solutions“ tables).
• Make sure that the temperature does not change during the calibration
procedure.
Calibration with Calibration Solution