Operation
1
The 765 Laboratory pH Meter
Package contents
After unpacking, please check the shipment for
completeness. The package should contain:
• 765 Laboratory pH Meter
• Power cord
• This instruction manual
Short device description
• The Model 765 is used for pH and ORP measure-
ments in the laboratory.
• Temperature compensation is automatic using a
Pt 100 or Pt 1000 temperature probe or manual by
entering the temperature.
• Buffer solutions of eleven different, selectable
buffer sets are provided for calibration. Buffer rec-
ognition is automatic via Calimatic
®
.
• The calibration timer alerts you when the preset
interval between two calibrations has expired.
• Sensoface
®
electrode monitoring checks the con-
nected electrode and provides information on the
electrode condition.
• Knick Fullcheck
®
device self-test checks the
device functions at the press of a key.
• Records of parameter setting, calibration and
diagnostics are particularly helpful for QM docu-
mentation to ISO 9000 and GLP. The records can
be directly output to a standard printer via the inte-
grated serial interface.