Operation
11
Typical setting procedure
The meter is factory set to calibration buffer set 01
(Mettler Toledo technical trade buffers). However,
you want to calibrate using the technical buffer
solutions to DIN 19 267, calibration buffer set 03.
Therefore, you must select buffer set 03 on the
Parameter level.
Press
to access the Parameter level.
Press
to select the EDIT menu.
Press
to open the EDIT menu.
Select buffer set parameter BUFF using
.
Now press
to change the buffer set. The right
display will flash.
Select buffer set 03, technical buffers to
DIN 19 267, using
.
par
enter