Operation
25
Electrode Data menu
The Electrode Data menu shows the values for
zero and slope of the electrode calculated during
the last calibration, the cal timer count, and the
electrode statistics.
Display of electrode zero determined during last
calibration.
Display of electrode slope determined during last
calibration.
Display of time in hours until the next calibration is
requested by the timer.
The electrode statistics include the electrode zero,
slope and response time of the last three calibra-
tions with corresponding date and time. In addition,
the data of the first calibration (see Pg. 15) per-
formed on this electrode are provided. This allows
you to evaluate electrode behavior over its whole
life span. Since the data are very comprehensive,
they should be printed out or read out via computer.
Date of last calibration
Date of second last calibration
Date of third last calibration
Date of first calibration