Installation and connection
Vacuum Pump System SC 920G
18
Translation of original Operating Instructions, English, KNF 121308-121311 07/18
6.2. Connection
1. Connect the vacuum chamber to the gas inlet's hose connec-
tor (Fig. 2/
7
, page 11, for hose inside diameter of 10 mm). For
this, vacuum tubing must be used.
2. Connect hose to high-performance condenser in order to
discharge gas exhaust (Fig. 2/
17
, page 11).
CAUTION
Danger of high performance condenser bursting.
The high-performance condenser is not pressure-
proof.
Do not reduce or regulate the quantity of gas at
the gas outlet, and do not install any compo-
nents that hinder the gas flow.
Safely discharge gas exhaust so that no gas can escape into
the ambient air.
Make sure that the high-performance condenser’s gas outlet is
not blocked (high-performance condenser is not pressure-
proof).
3. Attach coolant feed and return to high-performance condenser
19
and
20
, page 11).
Connect only the KNF coolant valve (see Chapter 11.2,
Accessories) to the coolant valve connection (Fig. 3/
2
, page
12). Consult with KNF before using any other valves.
4. If necessary: Connect inert gas supply to ventilation connec-
tion (Fig. 2/
6
, page 11). Observe the safety instructions in
chapter 3.
5. Insert the power cable plug into a properly installed shockproof
socket.