background image

Connecting the Pneumatic Pump K–1900 

 

 

Setting up the K-1900

 

 

 

Never run the Pneumatic Pump K–1900 without liquid in the pump head or 
in  the  piston  backflushing  compartment.  For  piston  backflushing,  follow 
the  instructions  below.  Operation  with-out  liquid  may  lead  to  damage  of 
the pump seals in short time.

 

 

Piston Backflushing

 

Backflushing  the  piston  removes  traces  of  salt  and  other  decontaminants  from 
the backside of the pistons. 

Especially  if  you  use  saline  solvents  or  buffer  solution  we  highly  recommend 
that  you  continuously  backflush  in  order  to  prevent  crystallization  which  can 
damage your piston seals. 

For  best  protection  and  error  free  long  time  operation  of  the  pump  head  it  is 
recommended  always  to  backflush  the  piston  with  an  appropriate  solvent 
(mostly water) according to the following SOP. 

 

 

SOP  1 

Preparing your K–1900 for piston backflushing

 

This SOP applies to the Pneumatic Pump K–1900. 

1. 

Push a 1/16" ID tube over both piston backflushing capillaries. 

 

 

 

 

 

 

 

 

 

Fig. 7 

Piston backflushing capillaries

 

 

2. 

Place one low end of the tubing in a flask. 

3. 

Fill the priming syringe with the backflushing liquid. 

4. 

Connect the syringe with the other tubing. 

5. 

Press liquid through the pump head, until it flows without any air bubbles 
into the container. 

6. 

Remove tubings from capillary opening. 

If  you  want  operation  with  continuous  backflushing  you  can  attach  two 
containers of rinsing liquid instead of the priming syringe. The containers should 
be positioned that one container is located higher than the other, thus ensuring 
liquid flow through the pump head without any assistance. 

When  not  backflushing  continuously,  we  recommend  to  connect  both  capillary 
openings with a tubing to prevent vaporization of solvents and drying out of the 
piston chamber. 

Summary of Contents for K-1900

Page 1: ...90 00 0 K K 1 19 95 50 0 M Ma an nu ua al l H Ha an nd db bu uc ch h V7062 02 2017 KNAUER Wissenschaftliche Ger te GmbH Hegauer Weg 38 D 14163 Berlin Germany Tel 49 0 30 809 727 0 Fax 49 0 30 801 50...

Page 2: ...17 Declaration of conformity 34 Inhalt Allgemeine Beschreibung der Pneumatischen Pumpe K 1900 18 Hinweise zum Gebrauch des Handbuchs 19 Konventionen in diesem Handbuch 19 SOP s in diesem Handbuch 19 T...

Page 3: ...pump head 0 1 499 9 ml min stainless steel up to 100 bar 1000 ml pump head 0 1 999 9 ml min stainless steel up to 50 bar The names of the pump heads have historical reasons You can also operate any o...

Page 4: ...rough all routine tasks during operation They can be used for documentation purposes and be copied applied signed and filed in order to document the performance of the instrument Please operate the in...

Page 5: ...ressure must under no circumstances ever exceed 4 2 bar Exceeding an external air pressure of 4 2 bar will lead to damage of pump head and pump mechanics This damage is not covered by guarantee Setup...

Page 6: ...air Fig 1 Front View of the Pneumatic Pump K 1900 1 8 2 1 2 2 2 3 2 4 1 7 1 1 1 6 1 3 1 4 1 5 1 2 1 Pump Head 2 Flushing Valve 1 1 Piston backflushing capillaries 2 1 Inlets to flushing valve 1 2 Elue...

Page 7: ...o the compressed air connector of the pump see Fig 2 To disconnect from the compressed air supply Hold the brass part of the coupler with one hand and pull the blue part of the coupler socket away fro...

Page 8: ...nts such as methanol and hexane rinse after using methanol with an intermediate solvent that is miscible with both solvents in all proportion e g isopropanol 100 250 and 500 ml Head Solvent tubings ar...

Page 9: ...according to the following SOP SOP 1 Preparing your K 1900 for piston backflushing This SOP applies to the Pneumatic Pump K 1900 1 Push a 1 16 ID tube over both piston backflushing capillaries Fig 7 P...

Page 10: ...ng device is connected to the solvent outlet Air Pressure Influences The resulting eluent pressure increases or decreases as a function of compressed air inlet pressure Eluent pressure will nearly be...

Page 11: ...want to use the pneumatic pump as a delivery pump with constant delivery volume you need to calibrate it Fig 11 shows typical calibration curves for two different columns with different pressure losse...

Page 12: ...onal spanner no 4 4 mm loosen just two diagonally opposed Head set screws Pos 1 3 Remove the screws 7 Carefully loosen the two remaining screws alternating from one to the other approx half a turn Thi...

Page 13: ...ng plate screws Pos 3 19 half a turn alternating from one to the other to avoid damaging the Retaining plate 3 Because the two screws are very tight it may be helpful to either clamp the pump head or...

Page 14: ...prings Pos 3 15 and the Guides for springs Pos 3 16 7 Put the Retaining plate Pos 3 18 over the Guides for springs Pos 3 16 with the conical bore of the plate directed outwards 8 Insert the two Retain...

Page 15: ...remove the check valves for cleaning from the pump head after loosening and removing the Inlet valve housing Pos 3 5 and Outlet valve housing Pos 3 6 and then disassemble them according to the follow...

Page 16: ...Set of fittings and adapters 1 8 1 16 including 1 1 8 1 16 transforming coupler 1 Fitting screw for capillary AD 1 8 M8x1 1 Ferrule for capillary AD 1 8 1 Fitting screw for capillary AD 1 16 1 Ferrul...

Page 17: ...sings 1 T peace fitting srcew with ferrule 1 left hand and 1 right hand connection tube A 1121 Spare Parts for Pump Head 250 ml Check valve 3 10 A 1122 Piston rod 3 8 3 17 A 1017 Set of Piston seals i...

Page 18: ...Edelstahl bis zu 100 bar 1000 ml Pumpenkopf Edelstahl bis zu 50 bar Die Bezeichnung der Pumpenk pfe hat historische Ursachen Sie k nnen die genannten Pumpenk pfe auch mit unserer Pr parativen HPLC Pu...

Page 19: ...lten schrittweise Anweisungen die den Anwender durch alle Aufgaben f hren Sie k nnen gleichfalls zu Dokumentationszwecken genutzt werden Sie k nnen kopiert angewendet unterzeichnet und abgelegt werden...

Page 20: ...4 2 bar f hrt zur Besch digung des Pumpenkopfes und der Pumpenmechanik Diese Sch den fallen nicht unter die Garantiegew hrleistung Inbetriebnahme der Pneumatischen Pumpe K 1900 Auspacken Alle KNAUER...

Page 21: ...Frontansicht mit Pumpenkopf und Druckaufnehmer 1 8 2 1 2 2 2 3 2 4 1 7 1 1 1 6 1 3 1 4 1 5 1 2 1 Pumpenkopf 2 Druckaufnehmer 1 1 Kolbenhintersp lungskapillare 2 1 Fl ssigkeitseinlass 1 2 Fl ssigkeitse...

Page 22: ...e Abb 3 Zum Trennen von der Druckluftquelle halten Sie die Schlaucholive der Druckkupplung mit einer Hand fest w hrend Sie den blauen Ring mit der anderen Hand von der Pumpe weg ziehen Druckluft leitu...

Page 23: ...smitteln unbegrenzt mischbar ist In den meisten F llen ist i Propanol geeignet 100 250 und 500 ml Kopf Die L sungsmittelschl uche werden entsprechend Abb 5 mit dem Pumpenkopf verbunden Je nach den kon...

Page 24: ...matische Pumpe K 1900 f r die Kolbenhintersp lung Diese SOP gilt f r die Pneumatische Pumpe K 1900 1 Schieben Sie zwei Schlauchst cke mit einem Innendurchmesser von 1 16 ber die Anschl sse f r die Kol...

Page 25: ...itsdruck erh ht sich oder sinkt in Abh ngigkeit vom Eingangsdruck der Pressluft Wird der Pressluftdruck konstant gehalten verbleibt der Fl ssigkeitsdruck ebenfalls praktisch konstant Die Pressluftdruc...

Page 26: ...r den 1000 ml Pumpenkopf Kalibrierung f r konstanten Fluss Um die pneumatische Pumpe f r die F rderung eines konstanten Volumens einzusetzen muss sie kalibriert werden Die Kalibrierungskurven haben f...

Page 27: ...6 L sen Sie mit einem 4 mm Imbusschl ssel zwei diagonal gegen ber liegende Schrauben der Pos 1 3 und entfernen Sie diese 7 L sen Sie die beiden verbleibenden Schrauben vorsichtig alternierend durch h...

Page 28: ...g eines Pumpenkopfes 1 Wenden Sie die SOP 4 Kolbenstangen ausbauen und pr fen an 2 L sen Sie die beiden Schrauben der Grundplatte Pos 3 19 vom Pumpenkopf Um Besch digungen der Grundplatte zu vermeiden...

Page 29: ...Teil 3 A 5 Schrauben Sie die Abstandsbolzen Pos 3 14 zun chst handfest ein ehe sie mit einem SW 10 Schl ssel angezogen werden 6 Setzen Sie die beiden Druckscheiben Pos 3 13 die Druckfedern Pos 3 15 un...

Page 30: ...nd schlie en sie nicht mehr korrekt Zum Reinigen k nnen Sie die Ventile durch L sen und Entfernen der Einlassventilhaltung Pos 3 5 und der Auslassventilhaltung Pos 3 6 aus dem Pumpenkopf ausbauen und...

Page 31: ...ngen Satz von Fittings und Adaptern 1 8 1 16 bestehend aus 1 1 8 1 16 bergangsst ck 1 Schlauchverschraubung f r Kapillare AD 1 8 M8x1 1 Schneidring f r Kapillare AD 1 8 1 Schlauchverschraubung f r Kap...

Page 32: ...end aus 2 Einlassventilgeh usen 1 T St ck Fittingschraube mit Schneidring 1 linker und 1 rechter Verbindungsschlauch A 1121 Pumpenkopfersatzteile 250 ml Kopf Kugelventil 3 10 A 1122 Kolbenstange 3 8 3...

Page 33: ...er Gew hrleistungsbedingungen Die werksseitige Gew hrleistung f r die WellChrom Pneumatische Pumpe K 1900 betr gt ein Jahr ab Versanddatum Unsachgem e Bedienung des Ger tes und Folgen einer Fremdeinwi...

Page 34: ...conformity The Declaration of Conformity is part of the delivery and accompanies the product as a separate document Konformit tserkl rung Die Konformit tserkl rung ist Bestandteil des Produktes und l...

Reviews: