Operating the valves and valve drives
DIP configuration of the electric valve drive
The DIP switches for configuration of the electric valve drive are
set and checked on the device bottom.
Prerequisite
Device is switched off.
Switch on the device after the new configuration so that the
new configuration becomes effective.
Principles of the setting for the DIP switch on the device
Fig. 38
Basic functions of the DIP switches
Fig. 39
Details of the setting of the DIP switches among the
different series of the valve drives
DIP
Function
Comment
1
Specify valve type
-
2
Specify valve position
-
3 and 4
Set the port OUT/BIN 3 on the
back of the device as input or
output
Set DIP
switches 3
and 4 always
opposite
DIP
I/0
V6
V12
V16
1
IN
360° valve
7P/1C
360° valve
13P/1C
not permitted
OUT
6P/2C
6P/3C
not permitted
360° valve
17P/1C
2
IN
BIN 0-2 at
360° valves
BIN 0-3
BIN 0-3
OUT
RS-232,
manual,
Pos. 1; [I]; [L]
RS-232,
manual,
Pos. 1; [I]; [L]
RS-232,
manual,
Pos. 1; [I]; [L]
3
IN
Output
OUT/BIN 3
Output
OUT/BIN 3
Output
OUT/BIN 3
4
OUT
3
OUT
not permitted
Input
OUT/BIN 3
Input
OUT/BIN 3
4
IN