![Knauer BlueShadow Series Manual Download Page 28](http://html1.mh-extra.com/html/knauer/blueshadow-series/blueshadow-series_manual_1979834028.webp)
29
Installation
Mounting plate for valves and column
holder
Valves can be attached on the side of housings of pumps, detec-
tors, valves, or multifunction modules (assistant) with a mount-
ing plate (order number G6725) and 4 screws.
Fig. 24
Mounting plate for valves and column holder
Connecting capillary with valve
and seal
Biconical seal
or
Dynaseal
You can use two seal types to connect the steel or PEEK capillary
with the valves – biconical – or
Dynaseal
.
Note:
For pressures
≥
50 bar you should always use
a
Dynaseal
.
Legend
A. Valve drive
B. Mounting plate
C. Electrical injection or
multiposition valve
D. Side holder for
valves
E. Column holder
A
B
C
D
E
A. Biconical seal
B. Cutting ring
C. Seal ring
Fig. 25
Biconical seal and Dynaseal
B
A
C