Installation
Fig. 21
Exploded view of an electrical valve
Mounting of the valve on the
valve drive
Delivery ex works
All valve drives are delivered with a ready-mounted injection or
multiposition valve. For repeat orders, the electrical valves are
delivered with included adapter plate.
Screw the adapter plate to the back side of the valve.
1. Attach the adapter plate the right way around. The openings
for the countersunk screws point to the front.
2. Fasten the adapter plate with two Phillips-head screws (M3).
Tighten the Phillips-head screw firmly.
Fig. 22
Back side of the electrical valve – adapter plate
for mounting
Legend
A. Axial plate
B. Axial needle bearing
C. Compensation plate
D. Disk spring
E. Rotor
F. Rotor seal
G. Seal ring
A
B
C
D
F
E
G
Legend
A. Back side of an elec-
trical valve
B. Markings for the
mounting
C. Opening for the
drive axle of the
valve drive
D. Adapter plate for
the mounting
E. Opening for coun-
tersunk screws
F. Phillips-head screws
A
B
C
D
E
F