Instructions for Use
Electrosurgical Unit ME 102
38
V. 3.1
11
Guidelines and Manufacturer’s Declaration Regarding
Electromagnetic Compatibility (EMC)
Guidance and manufacturer’s declaration according to IEC 60601-1-2,
clause 6.8.3.201 a) 3) Table 201: Electromagnetic emission
The device ME 102 is intended for use in an electromagnetic environment as specified below. The user
should assure that it is used in such an environment.
Emission test
Compliance level
Electromagnetic environment –
Guidance
RF emissions according to
EN 55011
Group 2
The device uses electromagnetic energy
exclusively for the performance of its
internal functions. This may cause
interference in nearby electronic equipment.
RF emissions according to
EN 55011
Class B
The limits of this class are kept only in
standby mode (without RF current
activation)!
Harmonic emissions according
to IEC 61000-3-2
Class A
The device is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Voltage fluctuations / flicker
emissions according to
IEC 61000-3-3
Complies